6416h GB/2021-01
29 / 86
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2.3
Commissioning and basic settings
Thoroughly rinse the ultrasonic bath oscillating tank with water before its
fi
rst use.
Note:
All outer surfaces (and also the inner walls of the oscillating tank) are covered with
an oily preservative in order to protect the surface during transport and storage.
This should be removed with a suitable cleanser before commissioning, see
Section 5.
Connect the TRISON Base to the mains switch for testing.
If the touchscreen lights up, the connection is correct.
Note:
If the Ethernet cable is not connected or if there is no network
connection, an error message will appear on the touchscreen after
approx. 13 sec. This will disappear automatically and the ultrasonic
bath will perform an automatic restart.
In order to initialise a network connection, the Ethernet cable must
be plugged in before turning on the device, or before the ultrasonic bath is turned o
ff
and on again.
It is recommended that a foil test be conducted as part of quality assurance prior to
the
fi
rst use.
This test is to be saved for later comparison, see Section 6.2.
Place the TRISON Rack or Twist in the ultrasonic oscillating tank. Place lid on top.
IMPORTANT!
Required basic settings must also be entered in the system prior to the
fi
rst use.
2.3.1 Basic settings
The required data are:
• Operating volume of the connected ultrasonic bath:
Only important for the duration of the degassing!
• Treatment time for the “MIS” program:
Only important for the duration of cleaning of MIS
instruments!
• Runtime for the "Robotics" program:
The duration of the soaking and cleaning phase for the
robotic instruments can be set.
Incorrect or false values shall be the responsibility of the user.