Caution!
The ultrasonic bath may not be opened.
Disconnect the ultrasonic bath from the mains before replacing the fuses.
6.4.3 Determining information on software version and report data
The system data is exclusively used for Service purposes. It must be communicated, when
required, to authorised technical personnel or to the manufacturer.
In
stand-by mode, press and hold the [
] key. A sub-menu will appear with diverse,
current temperature values for the ultrasonic bath (Service menu).
With a simple pressing of the [
] key, a further sub-menu with system data (e.g. serial
number and software version) may be opened.
With repeated, simple pressing of the [
] key, a third sub-menu will open with the
setting data for the HF generator (FH1).
The Service menu may be exited at any time by pressing the [START/STOP] key.
press and hold
T water
17.2 °C
T ch-liquid 25.2 °C
T tank
20.3 °C
water level
o.k.
* S O N O C O O L *
----------------------
12:00:00 17.0°C 50%
12:00:00
17.2°C
12:00:00
T water
17.2 °C
T ch-liquid 25.2 °C
T tank
20.3 °C
water level
o.k.
* S O N O C O O L *
----------------------
SN 3500.00000000.123
R 00.01-Dec 07 2009
ON time 000001:25:11
US time 000001:01:57
CH time 000001:14:20
sw/ar 00266/00000
* S O N O C O O L *
----------------------
SN 3500.00000000.123
R 00.01-Dec 07 2009
ON time 000001:25:11
US time 000001:01:57
CH time 000001:14:20
sw/ar 00266/00000
* S O N O C O O L *
----------------------
SN 09100676
R 02.04-2005-06-29
US time 000001:01:57
sw/ar 00167/00115
= US-Module (FH1) =
----------------------