51346-007 GB/2020-08
15 / 46
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
2.3
Start-up
Important!
During delivery, rubber stoppers are plugged into the tank's inlet and outlet so that any
residual water from the manufacturer's
fi
nal inspection does not leak into the internal
fl
uids
circuit during transportation.
Leave these stoppers in place for the time being!
Thoroughly wipe clean the ultrasonic bath's oscillating tank before its
fi
rst use.
Note:
In order to protect the surface during transport and storage, all outer surfaces (and also
the inner walls of the oscillating tank) are covered with an oily preservative. This should
be removed with a suitable cleanser before
fi
rst use, see Chapter 5.
Connect the ultrasonic bath up to the power supply (grounded socket).
To conduct a function test,
fi
ll and ventilate the oscillating tank in accordance with
chapter 4 . Brie
fl
y activate the ultrasound on the switched-on ultrasonic bath. A hissing
noise should be heard.
Check whether the cooling unit starts: An air current should be detectable on the rear
side of the outgoing air opening, the fan motor should be running audibly.
It is recommended that a foil test be conducted as part of quality assurance prior to the
fi
rst use, see chapter 6.1.
Next, empty out and switch o
ff
the ultrasonic bath in accordance with chapter 4.
Hang the beaker holder in the oscillating tank and place the lid on top.
Caution!
Safely store the rubber stoppers for later emptying or re
fi
lling of the tank, or for safe
transportation.