51346-007 GB/2020-08
23 / 46
1
electronic GmbH & Co. KG • Heinrichstraße 3-4 • 12207 Berlin • Deutschland • [email protected]
added accordingly - the liquid level may not drop below the inlet.
Note:
It is possible that liquid will need to be added several times until the internal tubing is
fully de-aired .
With inset beakers included, the
fi
ll level must be at or above the
fi
lling level mark since
a low
fi
ll level may cause damage to the ultrasonic bath .
Safely store the rubber stoppers for later emptying or re
fi
lling of the tank, or for safe
transportation.
Instructions –
fi
lling
• Do
not
fi
ll the oscillating tank with hot water (> 30 °C). The
fi
lling temperature should be
between 15 and 25 °C.
• At least drinking-quality water must be used to
fi
ll the oscillating tank.
• Only use distilled or deionised water without additives in inset beakers or insert tubs.
• Do not use any combustible
fl
uids (e.g. benzine, solvents) or chemicals that contain
chloride ions or that separate (some disinfectants, household cleaners, and dish
detergents) for sonication in the stainless steel tank.
• When working with aggressive preparations in inset beakers or insert tubs: Prevent
the contact
fl
uid or stainless steel surfaces from being sprayed. If needed, replace the
contact
fl
uid immediately, clean the surfaces and wipe dry.
• When using preparations, the safety instructions included in the product lea
fl
ets are to
be fundamentally adhered to.
4.2
Degassing the
fl
uid
Freshly-
fi
lled contact liquid or liquid that has remained in the oscillating tank for a long
period of time must be degassed prior to use. See also chapter 4.5.3.
Remove basket and other accessories from the oscillating tank.
Turn on the ultrasonic bath. The internal pump starts pumping, the cooling runs
automatically according to the set target temperature .
Note:
If cooling is not desired, it can be temporarily deactivated by adjusting the target
temperature, e.g. by setting the value to 30 °C. The ultrasound is switched on
separately (see chapter 3.4).
The
contact liquid must visibly and palpably flow into the tank through the (tank) inlet.
Set the degassing time (10 min) and activate the ultrasound, see chapter 3.5.