Fault messages on the display.
Critical errors are detected by the ultrasonic bath, signalled by the status LED, and shown
on the display with an error number . If the error cannot be recti
fi
ed, the ultrasonic bath must
be sent to Customer Service for inspection.
Overview of error messages issued :
Error
No.
Fault messages on
the display
Description / remedial action
002
-------------------------
=== Error 002 ===
-------------------------
T water
sensor error
Temperature sensor "contact liquid" is defective. The ultrasonic
bath interrupts all activities, the ultrasound can no longer be
started. The status LED blinks red.
Contact the manufacturer .
003
-------------------------
=== Error 003 ===
-------------------------
T ch-liquid
sensor error
Temperature sensor "coolant" is defective. The ultrasonic bath
interrupts all activities, the ultrasound can no longer be started.
The status LED blinks red.
Contact the manufacturer .
004 -------------------------
=== Error 004 ===
-------------------------
T tank
sensor error
Temperature sensor "oscillating tank" is defective. The ultrasonic
bath interrupts all activities, the ultrasound can no longer be
started. The status LED blinks red.
Contact the manufacturer .
010
Temperature low
The contact liquid temperature is too low.
No error. The contact liquid is additionally heated by the
ultrasound.
011
Temperature high
The contact liquid temperature is too high. The ultrasound cannot
be started and/or the ultrasonic bath turns o
ff
for safety reasons.
Wait until the cooling has su
ffi
ciently cooled the contact liquid
(tolerance range target tempe0.7 °C).
012
-------------------------
=== Error 012 ===
-------------------------
T ch-liquid 82.7°C
The coolant temperature is critically elevated. The ultrasonic bath
interrupts all activities until the temperature falls once again within
the normal range. The status LED blinks red.
Wait. Next, the last action conducted is automatically resumed.
013
-------------------------
=== Error 013 ===
-------------------------
excess temperature
T tank 34.8°C
1. Tank temperature is critical. The ultrasonic bath interrupts all
activities until the temperature falls once again within the normal
range. The status LED blinks red.
Wait.
2. After
fi
lling, an air bubble could have formed in the ultrasonic
bath, hindering the cooling circuit. Empty out and re
fi
ll under
observance of chapter 4.1.
014
-------------------------
=== Error 014 ===
-------------------------
level detect fail
Fill level is critically below the
fi
lling level mark or distilled water
without surfactant will be used as contact liquid, see chapter 4.1.
The ultrasonic bath interrupts all activities, the ultrasound can no
longer be started. The status LED blinks red.
Fill the contact liquid (water + concentrate in prescribed
quantities) up to the
fi
lling level mark and resume function with the
[START/STOP] key.
027
-------------------------
=== Error 027 ===
-------------------------
missing
= US module (FH1) =
An error message appears after the device is switched on.
The ultrasonic bath may continue to be used, but only with an
ultrasonic power of 100%.
Switch the ultrasonic bath o
ff
and back on. If the error appears
again, contact the manufacturer.