1586-006 en/2022-08
15/46
Preparation for operation
/4
4.3 Performing a function test
Requirement
• The device has adapted to the climatic conditions at the setup
location for at least 2 hours.
Procedure
1. Insert the mains plug for the device into an earthed mains
socket.
2. Press the on/off button to switch on the device.
3. Switch the ultrasound on briefly. Do this by pressing the start/
stop button. After 1 to 2 seconds, press the start/stop button
again to switch off the ultrasound.
Result
» A noticeable noise can be heard when the ultrasound is
switched on.
Contact our service department if you cannot hear any noise.
A foil test should be performed before the first use. This is used to
document the effectiveness of the ultrasound. See chapter
4.4 Rinsing out the oscillating tank
Thoroughly rinse the ultrasound bath’s oscillating tank with water
before first use.
In order to protect the surface during transport and storage, the
ultrasound bath is covered with an oily preservative. Before the
ultrasound bath is put into service, this preservative must be
removed with a suitable cleaning agent.