6/
Maintenance
32/46
1586-006 en/2022-08
7. The foil must be perforated, see
image. Else, we recommend
having the device checked by
the Service department at BAN-
DELIN electronic
GmbH & Co. KG: See chapter
8. Archive the foil with the test
date and serial number of the
ultrasonic bath. The foil test document template can also be
completed and archived.
9. Rinse the oscillating tank thoroughly to remove any detached
foil particles.
Suitable foil test frames can be ordered from BANDELIN electronic
GmbH & Co. KG. The foil test frames are designed for a wide range
of tank dimensions. Aluminium foil is also required to conduct the
test, but it is not included in the scope of delivery.
Type
Order No.
for
FT 4
3074
DL 102 H, DL 255 H
FT 6
3222
DL 156 BH
FT 14
3084
DL 510 H, DL 512 H, DL 514 BH
FT 40
3094
DL 1028 H
Содержание 7185.00159050.004
Страница 44: ...10 Accessories 44 46 1586 006 en 2022 08...