1586-006 en/2022-08
23/46
Operation
/5
Procedure
Hold down the button for setting the heating temperature and
press the on/off button to switch on the ultrasonic bath.
» As soon as the heating is switched on, the ultrasound is acti-
vated once per minute regardless of the temperature.
Information
– The device does not save the setting. The next time it is
switched off and back on again, the function for preventing
retardation of boiling will be active again from 60 °C.
5.9 Degassing the sonication fluid – DEGAS
Sonication fluid that has been freshly filled or left in the ultrasonic
oscillating tank for a long period must be degassed before use.
Degassing the sonication fluid increases the effect of the ultra-
sound. You can use the DEGAS function to perform rapid degassing
of the sonication fluid.
1. Press the start/stop button for at least 2 seconds.
» The ultrasound is switched on.
» The LEDs indicate the remaining degassing time.
2. To end degassing earlier, press the start/stop button.
Information
The ultrasound noise becomes quieter during degassing. This
means that the ultrasonic effect is increasing.