background image

BTL5-A/C/E/G1_-M/U_ _ _ _-K-SR 32/K_ _
Transductor de desplazamiento Micropulse – Forma constructiva de varilla

8

español

Resolución + Histéresis
= Reproducibilidad

Tension

0,3 mV

Corriente

0,6 µA

minimum

5 µm

Frecuencia de muestred:
Longitudes nominales

f

Standard

< 2000 mm

1 kHz

> 2000 mm

0,5 kHz

Desviación de linealidad:
Longitudes nominales
LN

< 500 mm

 > 500 mm

en µm

±100

 ±0,02 % FS

Coeficiente de temperatura
Salida de tensión:
[150 µV/K + (5 ppm/K * P * V/LN)] * 

T

Salida de corriente:
[0,6 µA/K + (10 ppm/K * P * I/LN)] * 

T

V = Margen de tensiones de salida

   en [V]

I

= Margen de intensidades de salida
   en [mA]

LN = Longitud nominal en [mm]

T = Diferencia de temperaturas en [K]

P = Posición del sensor de posición

   en [mm]

Resistencia a impactos 100 g/6 ms
según IEC 68-2-27 

1

Golpes permanentes

100 g/2 ms

según IEC 68-2-29 

1

Vibraciones  12 g, 10 hasta 2000 Hz
según IEC 68-2-6 

1

(Respetar/evitar la resonancias pro-
pias del tubo protector)
A prueba de presión hasta 600 bar
en el montaje en cilindros hidráulicos

1

Determinación individual según
norma de fábrica de Balluf

6

Características técnicas

Valores típicos para DC 24 V y 25 °C. Inmediatamente listo para funcionamiento, precisión total después de la fase de
calentamiento. Conjuntamente con el sensor de posición BTL-P-1013-4R, BTL-P-1013-4S o BTL-P-1012-4R:

6.1

Dimensiones, peso,
entorno

Longitudes nominales < 4000 mm
Dimensiones véase

 figura 3-2

Peso

aprox. 2 kg/m

Carcasa

Acero fino

Tubo protector Acero fino 1.4571
Diámetro

10,2 mm

Grosor de pared

  2 mm

Módulo de
elasticidad

en 200 kN/mm²

Fijación de la carcasa 6 tornillos
ISO 4762 M6 x 16 - A2-70
Temperatura de
empleo

–40 °C hasta +85 °C

Humedad < 90% sin condensación
Grado de protección según IEC 529
con conector montado:
Versión con conector

IP 67

Versión con cable

IP 68

(Homologación 5 bar/48 h)

6.2

Alimentación eléctrica

(externa)

Tensión estabilizada
BTL5-_1...

DC 20 hasta 28 V

Rizado

< 0,5 V

pp

Intensidad absorbida

< 150 mA

Intensidad pico
de conexión

< 3 A/0,5 ms

Protección contra inversión
de polaridad

Incorporada

Protección contra sobretensiones
Diodos protectores Transzorb
Resistencia a tensiones entre GND
(tierra) y carcasa

500 V

6.3

Salida

BTL5-A11...

Tensión salida

0...10 V/10...0 V

Intensidad de carga

< 5 mA

Rizado

< 5 mV

BTL5-G11...

Tensión salida  –10...10V/10... –10 V
Intensidad de carga

< 5 mA

Rizado

< 5 mV

BTL5-C1_...

Intensidad de carga   0...20/20...0 mA
Resistencia de carga  < 500 ohmios

BTL5-E1_...

Intensidad de carga  4...20/20...4 mA
Resistencia de carga  < 500 ohmios

6.4

Conexión con el módulo de
proceso

Interface analógica
Cable apantallado (longitud máxi-
ma, véase Conexiones),
Ø

 

6 hasta 8 mm

6.5

Alcance del suministro

Transductor de
desplazamiento

 figura 3-2

6.6

Sensor de posición

(debe pedirse por separado)

Sensor de posición
BTL-P-1013-4R, BTL-P-1013-4S,
BTL-P-1012-4R

Dimensiones

 figura 3-4

Peso

aprox.  10 g

Carcasa

Aluminio anodizado

Temperatura de
empleo

–40 °C hasta +85 °C

Sensor de posición

 

BTL5-P-4500-1

(electroimán)
Peso

aprox.  90 g

Carcasa

Plástico

Temperatura de
empleo

–40 °C hasta +60 °C

6.7

Accesorios

 (opcional)

Conector

 figura 4-3

N

ú

mer

o

 814 511 SP  

  Edici

ó

n 0108;  r

eservado el der

echo a intr

oducir modificaciones  

  Sustituye a edici

ó

n 9812.

7

Versiones (indicadas en la placa de características)

A

limentación eléctrica 1 = DC 24 V

Conexión eléctrica,

S32: con conector,

BTL5-A11-M0450-K-SR 32

K05: con cable, 5 m longitud

Versión de varilla,
Fijación: 6 tornillos ISO 4762 M6 x 16 - A2-70

        Longitud nominal (4 caracteres):

M = indicación métrica en mm
U = valor en 1/10 pulgada

 Interfaz analógica:
 Salida de tensión

A11 = 10 ... 0 V e 0 ... 10 V
G11 = 10 ... –10 V e –10 ... 10 V

 Salida de corriente

C10 = 0 ... 20 mA

E10 = 4 ... 20 mA

C17 = 20 ... 0 mA

E17 = 20 ... 4 mA

T

ransductor de desplaza-

miento Micr

opulse

Содержание BTL5-A11 series

Страница 1: ...ecniche Manuale d uso Descripci n t cnica Manual de instrucciones deutsch english fran ais italiano espa ol Balluff GmbH Schurwaldstra e 9 73765 Neuhausen a d F Deutschland Telefon 49 0 71 58 1 73 0 T...

Страница 2: ......

Страница 3: ...utsch Balluff GmbH Schurwaldstra e 9 73765 Neuhausen a d F Deutschland Telefon 49 0 71 58 1 73 0 Telefax 49 0 71 58 50 10 Servicehotline 49 0 71 58 1 73 3 70 E Mail balluff balluff de http www balluff...

Страница 4: ...64 Apparatus and Method for Auto matically Tuning the Gain of an Amplifier Emissionspr fungen Funkst rstrahlung EN 55011 Gruppe 1 Klasse A St rfestigkeitspr fungen Statische Elektrizit t ESD EN 61000...

Страница 5: ...ieht mit hoher Pr zision und Reproduzier barkeit innerhalb des als Nennl nge angegebenen Messbereichs Am Stabende befindet sich die D mpfungszone ein messtechnisch nicht nutzbarer Bereich der berfah r...

Страница 6: ...ungsbohrungen mit Zylinderschrauben ISO 4762 M6 x 16 A2 70 anzuschrauben Bild 3 2 Alle Schrauben m s sen mit 3 5 Nm festgedreht wer den Bei waagerechter Montage von Wegaufnehmern mit Nennl ngen gr er...

Страница 7: ...Netzoberwel len z B von Phasenanschnittsteue rungen f r die der Kabelschirm nur geringen Schutz bietet L nge des Kabels max 20 m 6 bis 8 mm L ngere Kabel sind einsetzbar wenn durch Aufbau Schirmung u...

Страница 8: ...rch falsche Verbindungen und berspannung besch digt werden Bevor Sie einschalten pr fen Sie deshalb die Anschl sse sorgf ltig 5 2 Null und Endpunkt einstellen Um den BTL Wegaufnehmer an die Erforderni...

Страница 9: ...t ver bundenen Komponenten ist regel m ig zu berpr fen und zu protokollieren 5 6 Funktionsst rung Wenn Anzeichen erkennbar sind dass das Wegmesssystem nicht ordnungsgem arbeitet ist es au er Betrieb z...

Страница 10: ...Kabelausf hrung IP 68 Typpr fung 5 bar 48 h 6 2 Stromversorgung extern Spannung stabilisiert BTL5 _1 DC 20 bis 28 V Restwelligkeit 0 5 Vss Stromaufnahme 150 mA Einschaltspitzenstrom 3 A 0 5 ms Verpol...

Страница 11: ...nglish Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a d F Germany Phone 49 0 71 58 1 73 0 Fax 49 0 71 58 50 10 Servicehotline 49 0 71 58 1 73 3 70 E Mail balluff balluff de http www balluff de http...

Страница 12: ...connection with this product US Patent 5 923 164 Apparatus and Method for Auto matically Tuning the Gain of an Amplifier Emission tests RF Emission EN 55011 Group 1 Class A Noise immunity tests Stati...

Страница 13: ...ar acteristic This process takes place with measuring high precision and repeatability within the stroke range defined as nominal stroke length At the rod end is a damping zone within which no reliabl...

Страница 14: ...ounting attach the transducer with all 6 cylinder head screws ISO 4762 M6 x 16 A2 70 Fig 3 2 All screws must be tightened to 3 5 Nm For horizontal mounting of trans ducer with stroke lengths greater t...

Страница 15: ...hase control devices against which the cable shield provides only limited protection Cable length max 20 m 6 to 8 mm Longer lengths may be used if construction shielding and routing are such that exte...

Страница 16: ...ble check your connections 5 2 Set null and span Should mounting conditions require it the factory set null and or end point for the magnet can be shifted by a total of max 15 of the full nominal stro...

Страница 17: ...its associated com ponents should be regularly checked and recorded 5 6 Fault conditions When there is evidence that the transducer system is not operating properly it should be taken out of service...

Страница 18: ...29 with connector attached connector version IP 67 cable version IP 68 type tested at 5 bar 48 h 6 2 Supply voltage external Regulated supply voltage BTL5 _1 DC 20 to 28 V Ripple 0 5 Vpp Current draw...

Страница 19: ...an ais Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a d F R F A T l phone 49 0 71 58 1 73 0 T l copieur 49 0 71 58 50 10 Servicehotline 49 0 71 58 1 73 3 70 E Mail balluff balluff de http www ballu...

Страница 20: ...of an Amplifier Contr les de l mission Rayonnement parasite EN 55011 groupe 1 classe A Contr les de la r sistance au brouillage Electricit statique ESD EN 61000 4 2 degr d intensit 3 Champs lectromag...

Страница 21: ...t lev e l int rieur de la plage de mesure correspondant la longueur nominale La zone d amortissement est situ e l extr mit de la tige et ne peut tre utilis e des fins de mesure Le capteur peut toutefo...

Страница 22: ...que ISO 4762 M6 16 A2 70 Fig 3 2 Toutes les vis doivent tre serr es au couple de 3 5 Nm En cas de montage horizontal de capteurs de d placement dont la longueur nominale d passe 500 mm il est recomman...

Страница 23: ...teur reliez le blindage au bo tier du connecteur multibroches BKS l int rieur de ce dernier fi gure 4 3 voir notice dans l em ballage du connecteur multibroches Dans le mod le raccordement fixe le bli...

Страница 24: ...nt la mise sous ten sion 5 2 R glage du point d origine et de la fin de course Pour adapter le BTL aux exigences propres l installation il est possi ble de modifier la position de d part et ou de fin...

Страница 25: ...f rents doit tre v rifi e r guli re ment et consign e 5 6 D faillance Lorsque des indices de dysfonc tionnement du syst me de mesure de d placement sont d cel s celui ci doit tre mis hors service et l...

Страница 26: ...marrage 3 A 0 5 ms D trompage incorpor Limiteur de tension Diodes de protection Transzorb Rigidit di lectrique GND contre le bo tier 500 V 6 3 Sorties BTL5 A11 Tension de sortie 0 10 10 0 V Courant d...

Страница 27: ...iano Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a d F Germania Telefono 49 0 71 58 1 73 0 Fax 49 0 71 58 50 10 Servicehotline 49 0 71 58 1 73 3 70 E Mail balluff balluff de http www balluff de ht...

Страница 28: ...gnati i seguenti brevetti US Patent 5 923 164 Apparatus and Method for Auto matically Tuning the Gain of an Amplifier Collaudi emissioni Irradiazione di disturbi radio EN 55011 Gruppo 1 Classe A Colla...

Страница 29: ...istica crescente o decrescente garan tita una precisione e riproducibilit elevata all interno del campo di misura Alla fine della barra si trova la zona di smorzamento non utilizzabile ai fini metrolo...

Страница 30: ...con su tutti e 6 i fori con viti cilindriche ISO 4762 M6 x 16 A2 70 Figura 3 2 Tutte le viti devono essere strette a 3 5 Nm In caso di montaggio orizzontale di trasduttori con campi di misura superior...

Страница 31: ...spina BKS figu ra 4 3 con l alloggiamento con nettore vedi le istruzioni compre se nel materiale di fornitura del connettore a spina Nella versione a cavo lo schermo del cavo collegato all alloggia me...

Страница 32: ...en dere il trasduttore controllare con cura le connessioni 5 2 Impostazione zero e fondo scala Per adattare il BTL alle esigenze possibile spostare il valore iniziale e o finale del datore di posizion...

Страница 33: ...di tutte le componenti ad esso connesse deve essere periodi camente verificato e protocollato 5 6 Difetti di funzionamento Qualora si individuino segnali che facciano presumere un funziona mento non...

Страница 34: ...entazione elettrica esterna Tensione stabilizzata BTL5 _1 DC 20 sino a 28 V Ondulazione residua 0 5 Vpp Assorbimento di corrente 150 mA Corrente massima di avviamento 3 A 0 5 ms Protezione contro inve...

Страница 35: ...es espa ol Balluff GmbH Schurwaldstrasse 9 73765 Neuhausen a d F Alemania Tel 49 0 71 58 1 73 0 Fax 49 0 71 58 50 10 L nea de servizio 49 0 71 58 173 370 E Mail balluff balluff de http www balluff de...

Страница 36: ...es US Patent 5 923 164 Apparatus and Method for Auto matically Tuning the Gain of an Amplifier Pruebas de emisiones Radiaci n con interferencias radiof nicas EN 55011 Grupo 1 clase A Pruebas de inmuni...

Страница 37: ...medida indicado como longitud nominal En el extremo final de la varilla se encuentra la zona de amortiguaci n una zona que no puede aprovecharse para medida y que puede rebasarse La conexi n el ctrica...

Страница 38: ...los 6 agujeros de fijaci n con tornillos cil ndricos ISO 4762 M6 x 16 A2 70 Figura 3 2 Todos los tornillos deben apretarse a 3 5 Nm En el montaje horizontal de trans ductores de desplazamiento con lo...

Страница 39: ...externos En el modelo enchufable la panta lla tiene que unirse a la carcasa del conector dentro del conector BKS figura 4 3 v anse las instruc ciones en el embalaje del conector En el modelo de cable...

Страница 40: ...efectuar la conexi n compruebe detenidamen te las conexiones 5 2 Ajuste del punto cero y del punto final Para adaptar el transductor de des plazamiento BTL a los requisitos la posici n inicial o fina...

Страница 41: ...omponentes asociados a ste debe verificarse peri dicamente y reflejarse en un protocolo 5 6 Anomal a funcional Si existen indicios de que el sistema de medici n de desplazamiento lineal no funciona de...

Страница 42: ...or IP 67 Versi n con cable IP 68 Homologaci n 5 bar 48 h 6 2 Alimentaci n el ctrica externa Tensi n estabilizada BTL5 _1 DC 20 hasta 28 V Rizado 0 5 Vpp Intensidad absorbida 150 mA Intensidad pico de...

Отзывы: