2 | ADROIT6000 Instruction Manual–English

2.3  Markings

The sensor will be marked with:

Refer to figure above and explanation below:
1. Model number. Refer to the product datasheet for identification.

Note: 

If the model number is followed by eight characters ‘

-

########’, refer to

manufacturer’s specification drawing ########.

2. Serial number.
3. Pressure units, pressure reference.
4. Supply voltage limits.

Note: 

Factory calibrated at 15

 

V.

5. Signal range and units.
6. Manufacturer’s name and address.
7. Country of assembly.

3. Installation and Operation

3.1  General Requirements

When the sensor is received, check the completeness in accordance with the supplied Calibration 

Certificate.
Read and understand all the related data before installing and using the sensor. This includes all 

local safety procedures and installation standards, this document and the product datasheet or 

specification drawing.
To identify the electrical and pressure connections, refer to the calibration certificate.
Ensure the sensor is mounted in a manner that avoids unwanted mechanical or thermal stress 

such as vibration, shock or excessive or rapid temperatures excursions.
A male pressure connector must not be sealed against the base of the thread.

The ambient temperature and the process media temperature to be measured must not exceed 

the ranges specified in the sensor specification.
For both gaseous and liquid media operating at low temperatures it is necessary to exclude 

crystallization or freezing components of the media from the working chambers and connecting 

pipelines.

ADROIT6###

PRESSURE SENSOR

ADROIT6###-T#-A#-CD-H0-P##

S/N ########

### TO ### ### #

Supply: ## TO ## V   ## mA

Output: ### TO ### ##

DRUCK LTD. LEICESTER, LE6 0FH, UK

MADE IN #####

[1]

[2]

[3]

[4]

[5]

[6]

[7]

9

8

Содержание Druck ADROIT6000

Страница 1: ...Druck com ADROIT6000 Pressure Sensor Instruction Manual nbn Austria GmbH...

Страница 2: ......

Страница 3: ...try of manufacture 2 Description 2 1 Purpose ADROIT6 pressure sensors hereinafter referred to as sensors are designed for continuous measurement and conversion of gauge atmospheric absolute and differ...

Страница 4: ...d using the sensor This includes all local safety procedures and installation standards this document and the product datasheet or specification drawing To identify the electrical and pressure connect...

Страница 5: ...recommended to install sensors in front of the shut off device if the medium to be measured is liquid If water hammer exists in the system it is recommended to install a hydraulic shock dampener Atta...

Страница 6: ...n use The sensors are calibrated during manufacture in the orientation pressure connector down and with the supply voltage at 15 V Supply voltage sensitivity is less than 0 005 FS V To adjust the Zero...

Страница 7: ...T6000 Instruction Manual 5 3 7 2 App Use Once the sensor is connected as shown follow the on screen instructions to adjust the ADROIT6 Please note that you will only be able to recalibrate up to 5 of...

Страница 8: ...uction Manual English 4 Key Documents A calibration certificate will have been shipped with your ADROIT6 pressure sensor The calibration certificate is printed exclusively in English Please find attac...

Страница 9: ...y Requirements Supply Voltage Output Signal Electrical Connection Wiring Details ve Supply ve Supply Case Room Temperature Calibration Pressure Measured Permitted bar a Deviation Deviation Performance...

Страница 10: ...Copyright 2020 Baker Hughes Company 8 ADROIT6000 Instruction Manual English...

Страница 11: ...Copyright 2020 Baker Hughes Company English ADROIT6000 Instruction Manual 9...

Страница 12: ...es com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 31 334678950 nl sensing sales bakerhughes com Moscow 7 915 3161487 aleksey khamov bakerhughes com UAE USA Abu Dha...

Отзывы: