Copyright 2020 Baker Hughes Company.
4 | ADROIT6000 Instruction Manual–English
3.5 Storage and Transport
Sensors in an individual package are to be transported by any kind of closed transport, in
accordance with the rules of transportation of goods acting on each mode of transport.
3.6 Returned Goods Procedure
If the sensor requires calibration or is unserviceable, return it to the nearest Druck Service Centre
Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information on either a RGA or RMA:
•
Product (e.g. ADROIT6200)
•
Serial number
•
Details of defect/work to be undertaken
•
Calibration traceability requirements
•
Operating conditions
3.7 Calibration and Adjustment
Calibration data can be collected from the sensor using standard reference pressure sources and
electrical measurement devices. Follow standard calibration procedures.
To optimize performance, it is recommended that calibration data is collected with the sensor:
a. In the same physical orientation as in use.
b. With the same supply voltage as in use.
The sensors are calibrated during manufacture in the orientation pressure connector down and
with the supply voltage at 15
V. Supply voltage sensitivity is less than 0.005%
FS/V.
To adjust the Zero only: a value of pressure applied and electrical signal output as close to, but
not below the lower range limit (LRL) of the sensor as practical must be recorded.
To adjust Zero and Span: values of pressure applied and electrical signal output, a Zero reading
as close to, but not below the LRL, and full scale reading as close to, but not above the upper
range limit (URL) as practical must be recorded.
Once this data is collected, the Zero setting or Zero and Span settings can be adjusted on the
ADROIT6### using the ADROIT
-
Interface in combination with the ADROIT6000 App.
3.7.1 Set-up
See
or Google Play Store for software download of the ADROIT6000 App.
It is recommended that calibration data is collected as a separate action in line with standard
calibration procedures as detailed in Section
To adjust the sensor connect the sensor as detailed below.
CAUTION
Adjustment by unauthorized sources may affect the warranty and
may not guarantee further performance.
ADROIT-Interface
ADROIT6###
+ve Supply
-ve Supply
USB
PC or Android™
Phone running
ADROIT6000 App
Signal (Voltage only) N/C