4 | ADROIT6000 Instruction Manual–English

3.5  Storage and Transport

Sensors in an individual package are to be transported by any kind of closed transport, in 

accordance with the rules of transportation of goods acting on each mode of transport.

3.6  Returned Goods Procedure

If the sensor requires calibration or is unserviceable, return it to the nearest Druck Service Centre 

listed at: 

Druck.com

Contact the Service Department to obtain a Return Goods/Material Authorization (RGA or RMA).
Provide the following information on either a RGA or RMA:

Product (e.g. ADROIT6200)

Serial number

Details of defect/work to be undertaken

Calibration traceability requirements

Operating conditions

3.7  Calibration and Adjustment

Calibration data can be collected from the sensor using standard reference pressure sources and 

electrical measurement devices. Follow standard calibration procedures.
To optimize performance, it is recommended that calibration data is collected with the sensor:
a. In the same physical orientation as in use.
b. With the same supply voltage as in use.
The sensors are calibrated during manufacture in the orientation pressure connector down and 

with the supply voltage at 15

 

V. Supply voltage sensitivity is less than 0.005%

 

FS/V.

To adjust the Zero only: a value of pressure applied and electrical signal output as close to, but 

not below the lower range limit (LRL) of the sensor as practical must be recorded.
To adjust Zero and Span: values of pressure applied and electrical signal output, a Zero reading 

as close to, but not below the LRL, and full scale reading as close to, but not above the upper 

range limit (URL) as practical must be recorded.
Once this data is collected, the Zero setting or Zero and Span settings can be adjusted on the 

ADROIT6### using the ADROIT

-

Interface in combination with the ADROIT6000 App.

3.7.1  Set-up

See 

Druck.com

 or Google Play Store for software download of the ADROIT6000 App.

It is recommended that calibration data is collected as a separate action in line with standard 

calibration procedures as detailed in Section

 

3.7.

To adjust the sensor connect the sensor as detailed below.

CAUTION  

Adjustment by unauthorized sources may affect the warranty and 

may not guarantee further performance.

ADROIT-Interface

ADROIT6###

+ve Supply

-ve Supply

USB

PC or Android™ 

Phone running 

ADROIT6000 App

Signal (Voltage only) N/C

Содержание Druck ADROIT6000

Страница 1: ...Druck com ADROIT6000 Pressure Sensor Instruction Manual nbn Austria GmbH...

Страница 2: ......

Страница 3: ...try of manufacture 2 Description 2 1 Purpose ADROIT6 pressure sensors hereinafter referred to as sensors are designed for continuous measurement and conversion of gauge atmospheric absolute and differ...

Страница 4: ...d using the sensor This includes all local safety procedures and installation standards this document and the product datasheet or specification drawing To identify the electrical and pressure connect...

Страница 5: ...recommended to install sensors in front of the shut off device if the medium to be measured is liquid If water hammer exists in the system it is recommended to install a hydraulic shock dampener Atta...

Страница 6: ...n use The sensors are calibrated during manufacture in the orientation pressure connector down and with the supply voltage at 15 V Supply voltage sensitivity is less than 0 005 FS V To adjust the Zero...

Страница 7: ...T6000 Instruction Manual 5 3 7 2 App Use Once the sensor is connected as shown follow the on screen instructions to adjust the ADROIT6 Please note that you will only be able to recalibrate up to 5 of...

Страница 8: ...uction Manual English 4 Key Documents A calibration certificate will have been shipped with your ADROIT6 pressure sensor The calibration certificate is printed exclusively in English Please find attac...

Страница 9: ...y Requirements Supply Voltage Output Signal Electrical Connection Wiring Details ve Supply ve Supply Case Room Temperature Calibration Pressure Measured Permitted bar a Deviation Deviation Performance...

Страница 10: ...Copyright 2020 Baker Hughes Company 8 ADROIT6000 Instruction Manual English...

Страница 11: ...Copyright 2020 Baker Hughes Company English ADROIT6000 Instruction Manual 9...

Страница 12: ...es com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 31 334678950 nl sensing sales bakerhughes com Moscow 7 915 3161487 aleksey khamov bakerhughes com UAE USA Abu Dha...

Отзывы: