background image

Copyright 2020 Baker Hughes Company.

English–ADROIT6000 Instruction Manual | 3

The materials used for the primary enclosure and pressure bearing surfaces are identified in the 

product data sheet or, if applicable, the specification drawing. Make sure that the materials are 

applicable for the installation.
Before using the sensor, remove the plastic/rubber protection cap from the pressure connector.

3.2  Safety Measures

The operation of sensors in systems where the pressure may exceed the overload values 

specified in the data sheet or customer-specific specification drawing is not allowed.
Connection and detachment of sensors from the mains supplying the pressure of the medium to 

be measured must be done after the shut-off valve is closed from the process and the pressure 

in the working chamber is made equal to atmospheric.
The connecting pipes should have a one-way slope (not less than 1:10) from the pressure 

collection point up to the sensor, if the medium to be measured is gas, and down to the sensor if 

the medium is liquid. If this is not possible, when measuring gas pressure at the lower points of 

the connecting lines, it is necessary to install sludge vessels, and when measuring the liquid 

pressure at the highest points, install gas collectors.
Selected devices for mounting sensors should be mounted on straight sections, at the maximum 

possible distance from pumps, locking devices, elbows, expansion joints and other hydraulic 

devices. It is especially not recommended to install sensors in front of the shut-off device if the 

medium to be measured is liquid. If water hammer exists in the system, it is recommended to 

install a hydraulic shock dampener.
Attach the sensor in a safe configuration that prevents unwanted stress (vibration, physical 

impact, shock, mechanical and thermal stresses). Do not install the sensor where it can be 

damaged by a material that causes corrosion. Provide additional protection for the sensor if it may 

be damaged in service.
When installing power supply and signal wiring, the possibility of condensate entering the sensor 

cable entry should be avoided.

3.3  Earthing (Grounding)

Connect the earth (ground) connections applicable to the installation.

3.4  Maintenance

The sensor contains no moving parts and requires a minimum of maintenance.
Do not try to do repairs to this sensor. Return the sensor to the manufacturer or an approved 

service agent.

3.4.1  Visual Inspection

Inspect the sensor for damage and corrosion. Any damage to the sensor must be assessed. If 

the housing is no longer sealed against water and/or dust, the sensor must be replaced.

3.4.2  Cleaning

Clean the case with a damp lint-free cloth and mild detergent.
If the product has been in contact with hazardous or toxic materials, obey all the applicable 

Control of Substances Hazardous to Health (COSHH) or Material Safety Data Sheet (MSDS) 

references and precautions when handling.

3.4.3  Metrological Characteristics

The metrological characteristics of the sensor correspond to the declared values during the inter-

testing interval as long as the consumer observes the storage, transportation and operation rules 

specified in this manual.

Содержание Druck ADROIT6000

Страница 1: ...Druck com ADROIT6000 Pressure Sensor Instruction Manual nbn Austria GmbH...

Страница 2: ......

Страница 3: ...try of manufacture 2 Description 2 1 Purpose ADROIT6 pressure sensors hereinafter referred to as sensors are designed for continuous measurement and conversion of gauge atmospheric absolute and differ...

Страница 4: ...d using the sensor This includes all local safety procedures and installation standards this document and the product datasheet or specification drawing To identify the electrical and pressure connect...

Страница 5: ...recommended to install sensors in front of the shut off device if the medium to be measured is liquid If water hammer exists in the system it is recommended to install a hydraulic shock dampener Atta...

Страница 6: ...n use The sensors are calibrated during manufacture in the orientation pressure connector down and with the supply voltage at 15 V Supply voltage sensitivity is less than 0 005 FS V To adjust the Zero...

Страница 7: ...T6000 Instruction Manual 5 3 7 2 App Use Once the sensor is connected as shown follow the on screen instructions to adjust the ADROIT6 Please note that you will only be able to recalibrate up to 5 of...

Страница 8: ...uction Manual English 4 Key Documents A calibration certificate will have been shipped with your ADROIT6 pressure sensor The calibration certificate is printed exclusively in English Please find attac...

Страница 9: ...y Requirements Supply Voltage Output Signal Electrical Connection Wiring Details ve Supply ve Supply Case Room Temperature Calibration Pressure Measured Permitted bar a Deviation Deviation Performance...

Страница 10: ...Copyright 2020 Baker Hughes Company 8 ADROIT6000 Instruction Manual English...

Страница 11: ...Copyright 2020 Baker Hughes Company English ADROIT6000 Instruction Manual 9...

Страница 12: ...es com Japan Netherlands Russia Tokyo 81 3 6890 4538 gesitj bakerhughes com Hoevelaken 31 334678950 nl sensing sales bakerhughes com Moscow 7 915 3161487 aleksey khamov bakerhughes com UAE USA Abu Dha...

Отзывы: