Copyright 2010 Baker Hughes Company.
28 | TERPS Instruction Manual–English
Reply:
000=> <Byte 0>,<Byte 1>, … ,<Byte 15><CR>
010=> <Byte 16>,<Byte 17>, … ,<Byte 31><CR>
…
0F0=> <Byte 240>,<Byte 241>, … ,<Byte 255><CR>
<CR>
Send <CR> to continue<CR>
(Waits 20 seconds or for CR)
100=> <Byte 256>,<Byte 257>, ... ,<Byte 271><CR>
...
1F0=> <Byte 496>,<Byte 497>, … ,<Byte 511><CR>
Send <CR> to continue<CR>
(Waits 20 seconds or for CR)
6.9 Error Messages
Error messages are generated from the following sources:
6.9.1 Incorrect User Command
In a command string, the transducer accepts correct commands and rejects incorrect commands.
An incorrect command causes the transducer to send an error message related to the command.
The form of the error message:
Reply:
ERROR nn
Where nn is a decimal number, detailed in Table
6.9.2 Transducer Fault
A transducer fault sends an error message blocking the transmission of pressure data and
sending the fault message in place of data.
An applied pressure, 5% of span out of the calibrated range, causes:
Reply:
*Over Pressure*
Reply:
*Under Pressure*
A damaged transducer or no frequency output from the transducer causes:
Reply:
**** NO RPT ****