Copyright 2010 Baker Hughes Company.
English–TERPS Instruction Manual | 29
Appendix
A. Fault Finding
Table
A
-
1: Fault Finding
Error Message Probable cause
Action
!001 Buf
Overflow
More than 30 characters sent without an
end of the line character.
Send 30 delete characters followed by an
end of line to clear the buffer.
!002 EEPROM
Error
Failure of non-volatile memory.
Return to Druck for repair or replacement.
!004 Bad
Command
Command character not recognized.
Check command string sent. Check the
serial interface connection.
!005 Bad Char
Illegal character sent in a command
string.
Check command string sent is valid.
Check the serial interface connection.
!006 Bad
Param(s)
Invalid parameters sent in command
string - out of range, too long, wrong type
etc.
Check command string sent.
!008 Bad Format Invalid formatting character sent in
command string or command is
interactive.
Check command string sent requires a
valid formatting character.
!009 Miss’g
Param
At least one of the required parameters
missing from command string.
Check command string sent.
!010 Invalid PIN Incorrect PIN for the command sent.
Check command string sent.
!011 Bad Value
Out of range value sent to the transducer. Send in range value.
!012 Bad BUS
Cmd
Command sent to transducer in Network
mode.
Send this command with transducer in
Direct mode.
!013 Cal Error
Corrupt calibration data detected,
transducer pressure measurement
unreliable.
Return to Druck for recalibration.
!014 Press
Range
Corrupt calibration data detected,
pressure range not known, transducer
pressure measurement unreliable.
Return to Druck for recalibration.
!015 Under Press Pressure applied below calibrated limit.
Check for system fault where transducer is
installed.
!016 Over Press Pressure applied above calibrated limit. Check for system fault where transducer is
installed.
!017 Bad Global Command sent to the transducer as a
global 0: command not valid as a global
command in network mode.
Send the command to each transducer in
turn, using correct address in place of
global 0.
!018 Bad
Response
Invalid value entered in response to a
prompt normally happens with inactive
commands.
Repeat command with valid values in
response to each prompt.
!019 Timed Out
Key not pressed in the specified time
period (usually 20 seconds) normally
happens with inactive commands.
Repeat command, entering response in
less than 20 seconds between each
character and send carriage return within
the specified time.
!020 No
Frequency
Change of measurement speed, see
Wait for 2 seconds, if fault remains return to
Druck for repair or replacement.
Содержание Druck 8000 Series
Страница 1: ...Druck com 8000 8100 8200 8300 Series Trench Etched Resonant Pressure Sensor Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2010 Baker Hughes Company iv TERPS Instruction Manual English...
Страница 42: ...Copyright 2010 Baker Hughes Company 34 TERPS Instruction Manual English Appendix B RPS EEPROM Format...
Страница 43: ......