Query:
<SPACE>*O,?<CR>
Reply:
Baud Rate = <Baud Rate><CR>
Parity = <Parity - Ignore/None/Odd/Even><CR>
Data Bits = <Char Length><CR>
Stop Bits = <Stop Bits><CR>
Handshake = <Software Handshaking - Yes/No><CR>
Term Chars = <Number of Termination Characters><CR>
Command:
<SPACE>*O<CR>
Reply:
(Interactive)
Reply:
Enter PIN => (enter PIN)
Reply:
Select Baud rate (0 - 6) => (enter selection)
Reply:
Enter parity ([N]one/[I]gnore/[E]ven/[O]dd) => (Enter parity)
Reply:
Enter number of data bits (7/8) => (Enter number of data bits)
Reply:
Enter number of stop bits (1/2) => (Enter number of stop bits)
Reply:
Software Handshaking? (Y/N) => (Enter Y or N)
Reply:
Number of termination characters (1 = CR, 2 = CRLF) => (Enter number
of termination characters)
Reply:
Accept these values? (Y/N) => (Enter Y or N)
(Y answered) “EEPROM updated”<CR>
(N answered) “Current settings left unchanged”<CR>
Result:
These settings take effect after switching off then on the power supply for
the transducer.
6.7.5 P - Change PIN
This command sets the PIN of the transducer that must be used to access the PIN protected
settings. To change this setting the current PIN must be used and can be integers between 0 and
999.
Command:
<SPACE>P, <Current PIN>,<New PIN><CR>
Reply:
n/a
Result:
The PIN of the transducer changes to new PIN.
0
-
19200 bps.<CR>
1
-
9600 bps.<CR>
2
-
4800 bps.<CR>
3
-
2400 bps.<CR>
4
-
1200 bps.<CR>
5
-
600 bps.<CR>
6
-
300 bps.<CR>
where: <Current PIN> is the PIN of the transducer.
<New PIN>
is the replacement PIN.
Содержание Druck 8000 Series
Страница 1: ...Druck com 8000 8100 8200 8300 Series Trench Etched Resonant Pressure Sensor Instruction Manual...
Страница 2: ......
Страница 6: ...Copyright 2010 Baker Hughes Company iv TERPS Instruction Manual English...
Страница 42: ...Copyright 2010 Baker Hughes Company 34 TERPS Instruction Manual English Appendix B RPS EEPROM Format...
Страница 43: ......