TBN581 (ORP/pION) TRANSMITTER
CALIBRATION
I - 4
WBPEEUI110501A0
CALIBRATION
This section details the calibration functions of the STT03E terminal while
interfacing a TBN581 ORP/pION transmitter. There are four types of cali-
bration functions:
•
Process calibration.
•
Bench calibration (ORP/pION).
•
Bench calibration (mV).
•
D-to-A adjust.
When calibrating, a device must be connected to the STT03E transmitter.
When using an IMFBS01 module (field bus), up to fifteen transmitters can
be connected to the bus. Select the transmitter by using
.
NOTE:
To change calibration or configuration parameters of a smart trans-
mitter that is connected to an IMFBS01 module, the transmitter must be
taken off-line. This is done at the INFI 90 OPEN console.
After selecting a device, the
READY
screen appears. All calibration func-
tions can be performed on the selected device. Refer to Figure
for an
overview of the calibration functions.
Process Calibration
This procedure applies to TBN581 Transmitters configured for either
ANA-
LOG
or
DIGITAL
. Any difference between the two are noted in the
Com-
ments
column.
Bench Calibration (ORP/pION)
This procedure applies to TBN581 Transmitters configured for either
ANA-
LOG
or
DIGITAL
. Any difference between the two are noted in the
Com-
ments
column.
Select
YES
to store the configuration.
The Configuration ID tag name just configured will be in
the upper left corner if the configuration is saved. It
becomes the working configuration.
NOTE:
Connect the terminal to a device and send the
configuration (using
) to a device before
calibrating.
Key
Display
Comments
ENTER
STORE THIS
CONFIGURATION
NO
→
YES
ENTER
CONFIG. ID TAG
READY
SEND CONFIG
SELECT DEVICE