TBN580 TRANSMITTER
CALIBRATION
WBPEEUI110501A0
H - 11
Use the arrow keys to adjust the 4 mA signal.
NOTE:
When increasing or decreasing the mA signal, the
increments of change increase with successive depressions
until the maximum level of change is reached. By chang-
ing direction you will return to the smallest increment of
change. This adjustment technique speeds up the adjust-
ment process without affecting fine adjustment.
Use the arrow keys to adjust the 20 mA signal.
Key
Display
Comments
ENTER
CALIBRATION
PROCESS CAL
BENCH CAL pH
BENCH CAL TEMP
→
D-TO-A ADJUST
ENTER
ADJUST TO 4 mA
THEN HIT ENTER
ENTER
ADJUST TO 20 mA
THEN HIT ENTER
ENTER
CONFIG. ID TAG
READY