TBN480 CONDUCTIVITY TRANSMITTER
CALIBRATION
G - 8
WBPEEUI110501A0
Editing the Calibration Constants
These functions are for recording or modifying slope or offset calibration
constants.
EDIT CONDUCTIVITY
This procedure steps through the modification of the conductivity.
Key
Display
Comments
Select
YES
.
Select
EDIT CAL.
Select
CONDUCTIVITY
.
To view values press
twice.
To change the slope enter the correct number and press
.
To change the offset, enter the correct number and press
.
CALI-
BRATE
OUTPUT WILL BE
AFFECTED
PROCEED?
NO
→
YES
ENTER
CALIBRATION
PROCESS CAL
→
EDIT CAL
RESET CAL
ENTER
EDIT CAL DATA
→
CONDUCTIVITY
TEMPERATURE
ENTER
CONDUCTIVITY
PROCESS CAL
SLP: n.nnnn
OFF: n.nnnn
ENTER
ENTER
ENTER
ENTER
WORKING
CALIBRATION
COMPLETE