![Bacharach Portable Combustion Analyzer Скачать руководство пользователя страница 52](http://html1.mh-extra.com/html/bacharach/portable-combustion-analyzer/portable-combustion-analyzer_instruction_3443059052.webp)
Calibration
5.4 Calibrate TS-Zero
Material Required:
•
Thermocouple Simulator (K-type) Range: 0 to 600 °F
Accuracy: ±0.5 °F
Procedure:
PCA
1. With the analyzer turned off, first plug the simulator’s K-type connec-
tor into the T-STACK jack (Figure 3-2); then turn on the analyzer and
wait for its warm-up cycle to complete.
2. Enter the Calibration Menu Screen per Section 5.3; then choose
TS-ZERO to zero the analyzer’s stack temperature channel.
3. Adjust the simulator to 32 °F (0 °C).*
4. Wait until the MEASURED reading on the screen stabilizes. Then use
the
st
and
ENTER
keys to enter an APPLIED value that equals 32 °F
(0 °C).
Typical Calibrate TS-Zero Screen During Calibration Procedure:
CALIBRATE TS-ZERO
MEASURED 37.0°F
APPLIED 0032.0°F
4. Press the
RUN
key to calibrate the analyzer’s MEASURED value to that
of the APPLIED value. At this time the Calibrate Menu Screen is
redisplayed.
* The TS-Zero calibration range is 32 – 41 °F (0 – 5 °C). Any attempt to
calibrate outside this range will cause the analyzer to display the
message BAD CALIBRATION ENTRY.
5-4
Instruction 0024-9219
Содержание Portable Combustion Analyzer
Страница 12: ...Technical Characteristics NOTES PCA 2 4 Instruction 0024 9219...
Страница 60: ...Calibration PCA NOTES 5 12 Instruction 0024 9219...
Страница 81: ...PCA Parts Service Figure 8 1 Replacement Parts 1 of 3 Instruction 0024 9219 8 3...
Страница 83: ...PCA Parts Service Figure 8 1 Replacement Parts 3 of 3 Instruction 0024 9219 8 5...
Страница 90: ...Appendix B PCA NOTES B 2 Instruction 0024 9219...