PCA
4.3 Performing a Combustion Test
Operation
Important! Large rapid changes in the temperature of the
analyzer can affect its accuracy. This is important to know if
the analyzer is stored in a cold place (such as an unheated
vehicle in the winter) and then taken into a warm furnace
area. For the most accurate test results, allow the analyzer to
warm up to room temperature before use (about 10 minutes).
4.3.1 Analyzer Turn On and Warm Up
Important! Be sure the probe is at room temperature before
performing the following steps.
1. Make sure that the analyzer is properly set up per Section 3.0.
2. Place probe in an area of fresh, ambient air; then press the analyzer’s
ON
/
OFF
key.
3. Wait for the analyzer to countdown through its 60 second warm-up
period; then perform one of the following:
• If no errors were detected during warm-up,
the Combustion
Test Screen will be displayed. Skip Step 4, and go to Section 4.3.2.
• If an error was detected during warm-up,
proceed with Step 4.
4. If one or more errors were detected by the analyzer’s microprocessor
during warm-up, these errors will be displayed at the bottom of the
Sensor Status Screen (Section 4.6). Address any problems now per
Section 7.2; then repeat this procedure starting with Step 1.
NOTE: If the error detected is not critical to your test, the
instrument can still perform any test not using the function
disabled by the error.
Instruction 0024-9219
4-3
Содержание Portable Combustion Analyzer
Страница 12: ...Technical Characteristics NOTES PCA 2 4 Instruction 0024 9219...
Страница 60: ...Calibration PCA NOTES 5 12 Instruction 0024 9219...
Страница 81: ...PCA Parts Service Figure 8 1 Replacement Parts 1 of 3 Instruction 0024 9219 8 3...
Страница 83: ...PCA Parts Service Figure 8 1 Replacement Parts 3 of 3 Instruction 0024 9219 8 5...
Страница 90: ...Appendix B PCA NOTES B 2 Instruction 0024 9219...