Instruction 24-9448
7-1
Troubleshooting
7 Troubleshooting
7.1 Analyzer Repair
It s recommended that feld repar of the PCA 2 be lmted to:
Checks of prnted crcut board connectors
Replacng the probe assembly
Replacng the flter element n the water trap / flter assembly
Replacng sensors
Replacng ether the gas or purge pump
Informaton on how to perform these repars s provded n Secton 6.
All other repars should be performed by an authorzed Bacharach Servce
Center (refer to Secton 8.3). Any repars performed by an
unauthor-
ized
servce organzaton wll vod the analyzer’s warranty and release
Bacharach, Inc. of any mpled or wrtten product lablty.
7.2 Error Symbols
Error symbols are shown n the data felds of the Combuston Test screens
and on the prntout to ndcate the followng:
“
- - -
” Three hyphens or dashes ndcates that the
calculated
data can-
not be dsplayed because the
measured
data necessary to make
the calculaton s out of range. For calculatons to occur, the oxy-
gen level must be below 16% and the stack temperature must be
below 2,000 °F (1,093 °C).
“
∗ ∗ ∗
”
Three stars appear n the data feld of sensors that are not
nstalled, defectve, or found to be n overrange durng warm-
up, and also n the data felds of that sensor’s related calculated
values. For example, f the data feld of ether the NO or NO
2
sensor contans three stars, then the NOx data feld wll also
contan three stars.
"X X X" Indcates sensor overrange.
•
•
•
•
•
Содержание PCA 2 24-9448
Страница 6: ...Instruction 24 9448 iv Contents Notes ...
Страница 20: ...Instruction 24 9448 2 4 Specifications Notes ...
Страница 66: ...Instruction 24 9448 4 32 Operation Notes ...
Страница 95: ...Instruction 24 9448 6 15 Maintenance Notes ...
Страница 99: ...Instruction 24 9448 7 4 Troubleshooting Notes ...
Страница 103: ...Instruction 24 9448 8 4 Parts Service 4 10 7 6 2 1 9 18 17 13 3 5 8 19 12 14 20 21 11 19 Figure 8 1 Parts 2 of 2 ...
Страница 107: ...Notes ...