MVR-300™ Refrigerant Gas Detector
28
P/N: 6203-9000 Version 1
5.3.3
Fault Conditions
Bit
Mask
(HEX)
Fault
Description
Fault
State
Pri-
ority
Clearing Action
Required
Actions During
Fault
All bits
cleared
No fault
0x0001
Software fault
Firmware error (e.g.,
unexpected switch
state)
Fatal
fault
1
Acknowledge (reset
firmware)
Wait for switch 2
hold
0x0002
Sensor out
Cannot detect sensor
Critical
fault
2
Sensor Detected
Check sensor
0x0004
Input voltage
fault
Power supply voltage
out of range.
Critical
fault
2
Input voltage within
specification
Check Input
voltage
0x0008
Memory fault
Error in
reading/writing to
RAM, flash or internal
(PIC) EEPROM
Critical
fault
2
Memory test
passed
Test memory
0x0010
DAC fault
Error updating DAC
value
Non
critical
fault
4
DAC can be written
to
Write to DAC
0x0020
Stuck
magnetic
switch
Magnetic switch
activated for > 1
minute
Non
critical
fault
4
Switch released
Check switch
state
0x0040
Negative gas
concentration
fault
Sensor output has
drifted too negative
Nega-
tive gas
fault
3
Gas concentration
exceeds negative
gas limit
Check gas
concentration
0x0080
Invalid
calibration
Error in calibration
configuration
Critical
fault
2
Load valid
calibration
Read from
external EEPROM
0x0100
Zero
calibration
failure
Zero calibration failed
Calibra-
tion
failure
5
Acknowledgement
of failed calibration
Wait for switch 2
hold
0x0200
Sensor read
EEPROM fault
Error in reading from
external EEPROM
Critical
fault
2
External EEPROM
read success
Read from
external EEPROM
0x0400
Sensor write
EEPROM fault
Error in reading
writing to external
EEPROM
Critical
fault
2
External EEPROM
write success
Wait for write
0x0800
Sensor
configuration
fault
Error in external
EEPROM data
Critical
fault
2
External EEPROM
write valid data
Wait for write
0x1000
Span
calibration
failure
Span calibration
failed
Calibra-
tion
failure
5
Acknowledgement
of failed calibration
Wait for switch 2
hold
0x2000
System read
EEPROM fault
Error in reading from
internal EEPROM
Critical
fault
2
Internal EEPROM
read/write success
Read from internal
EEPROM
0x4000
System write
EEPROM fault
Error in reading
writing to internal
EEPROM
Critical
fault
2
Internal EEPROM
read/write success
Wait for write
0x1000
System
configuration
fault
Error in internal
EEPROM data
Critical
fault
2
Internal EEPROM
write valid data
Wait for write
www.
.com
1.800.561.8187