background image

H-SEM 

Operation 

Instruction 3015-4584 

 

5

 

3 Operation 

3.1  Front Panel Display and Controls 

 
 
 
 
 
 
 
 
 
 
 
 
 
 

* Enter the Control/Setup Menu Mode by pressing 

the LEFT and RIGHT arrow buttons at the same time. 

 

3.2 Inspection 

The H-SEM unit has been thoroughly inspected and tested prior to shipment from the factory. 
Nevertheless, it is recommended that the monitor be re-checked prior to use. Inspect the outside of the 

enclosure to make sure there are no obvious signs of shipping damage. Open the enclosure and inspect the 
interior of the monitor for loose components that may have become dislodged during shipment. If damage 

is discovered, please contact your supplier for assistance. 

3.3 General Operation 

WARNING:

  

Operate the monitor in its horizontal position (handle facing upwards). Tilting the 

monitor to a vertical position while in operation may cause inaccurate measurements to occur.

 

To turn ON the monitor, first lift up the shield located in front of the handle, and then press the red power 

ON/OFF toggle switch.  

Once the monitor has been powered ON, allow it to warm up for 15 minutes; after which, press the 

ENTER

 button to fill the purge-air bag with

 clean, fresh air that is of the same temperature and 

relative humidity as the area being checked for gas. DO NOT fill the purge-air bag in an area 
that is contaminated with refrigerant gas.

  

After the purge-air bag has been filled, the monitor will automatically start making measurements in the 
area being sampled. The results of those measurements are displayed on the front panel display.  

MEASURE

 indicates when the monitor is actively measuring gas. The 

screen to the left shows that currently 10 ppm of R-134A refrigerant gas is 
being detected, and that a peak measurement of 20 ppm has been made. The 

operator can reset the peak value to zero by pressing the 

ENTER

 button. 

The measurement cycle runs for 4 minutes. A log of up to 200 previous 
measurements can be viewed using the 

PPM LOG 

function (Page 8). 

MEASURE  00020pk 
   10ppm R134A 

DISPLAY 
SCREEN 

ENTER 
• Press to zero peak reading 
• When in the Control/Setup 

Menu Mode*, press to save a 
displayed value

KEYPAD 
When in the Control/Setup Menu 
Mode*, use these buttons:  
• To move the arrow (>) on the 

display screen to the desired 
function 

• Scroll through data 
• Change a function’s value 

MONITOR ON 
LIGHT (Green) 
Flashes during 
warm-up 

SYSTEM FAULT 
LIGHT (Yellow) 

ALARM LIGHT (Red) 

SILENCE / ESC 
Press this button: 
• To temporarily silence the 

audible alarm 

• Return to the previous 

screen without saving data 

Power ON/OFF 
Toggle Switch 

Содержание H-SEM

Страница 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Страница 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Страница 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Страница 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Страница 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Страница 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Страница 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Страница 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Страница 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Страница 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Страница 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Страница 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Страница 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Страница 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Страница 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Страница 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Страница 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Страница 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Страница 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Страница 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Страница 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Страница 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Страница 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Страница 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Отзывы: