background image

Operation 

H-SEM 

Instruction 3015-4584

 

From the 

Data Display

 screen, press both

 

the

 Keypad

 

Left

 and 

Right

 buttons at the same time to 

display the first 

Function Menu

 screen. Next, use the 

Keypad

 buttons to move the arrow (>) through 

the menu screens until the arrow is next to the desired function, and then press the 

ENTER

 button to 

select that function.  

Once a function has been selected, use the 

Keypad

 to scroll through the displayed data or to change a 

parameter associated with that function. Press 

ENTER

 to save any newly entered parameters. Press the 

ESC 

button to return to the previous screen without saving. 

Note that if no buttons are pressed within 90 seconds after selecting a function, the unit returns to the 

Data Display

 screen. 

FILLBAG

 – Press 

ENTER

 to manually fill the purge-air bag. 

WARNING:

 

Fill the purge-air bag in a clean, fresh-air environment that is 

at the same temperature and humidity as the area being checked for gas. 

DO NOT fill the bag in an area that is contaminated with refrigerant gas.

 

EMTYBAG

 – Press 

ENTER

 to manually empty the purge-air bag. Use 

this function to drain the bag if it becomes contaminated with gas. 

ALRMLVL

 – 

Sets the detected gas ppm level at which the unit goes into 

an alarm state, as indicated by the front panel 

ALARM

 light turning ON. 

 
AUDALRM

 – 

Allows the monitor’s internal audible alarm to be 

associated with any function of the monitoring system. Factory default is 

OFF. Use the 

Keypad

 to select the desired audible alarm function, and 

then press 

ENTER

 to save that value and return to the previous screen.

  

Audible Alarm Settings:

 OFF, ALARM,  ALARM/FAULT 

PPM LOG

 – Contains records of the last 200 measurements. Each 

record shows the measurement’s date, time, and ppm level. 
Measurements are logged at an interval determined by the 

LOG INT 

function (Page 

1

8). 

Use the 

Keypad

 

Up

 and 

Down

 buttons to change the record number by 

a factor of 1. Use the 

Right

 and 

Left

 buttons to change the record 

number by a factor of 10. Press 

ESC 

to return to the previous screen. 

To the right, record #025 shows that a level of 100 ppm was measured on 

06/07/05 at 15:35. 
The PPM Log can be cleared as described under the heading 

Clearing the 

PPM Log & Faults Log

 on Page 12. 

LOG INT

 – Sets the interval at which measurements are logged to 

memory from 1 to 9999 minutes. Factory default is 1 minute. The logged 
measurements can be viewed using the 

PPM LOG 

function (Page 8). Use 

the 

Keypad

 to enter the desired value, and then press 

ENTER

 to save 

that value and return to the previous screen. 

DIAG

 – Enters the diagnostic function. Refer to the heading 

  Working 

with the DIAG 

Function on Page 12. 

 

 
 

#025    100PPM @
06/07/05  15:35

FILLING BAG... 

DRAINING BAG... 

AUDIBLE ALARM IS

 OFF 

EDIT ALARM LEVEL

    0100 ppm

LOG INTERVAL IS

    0001 min

4.2159v   <0000>
29.05cD 14.58psi

Содержание H-SEM

Страница 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Страница 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Страница 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Страница 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Страница 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Страница 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Страница 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Страница 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Страница 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Страница 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Страница 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Страница 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Страница 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Страница 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Страница 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Страница 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Страница 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Страница 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Страница 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Страница 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Страница 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Страница 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Страница 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Страница 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Отзывы: