background image

H-SEM 

Operation 

Instruction 3015-4584 

 

11

 

3.8.3 Viewing the Faults Log 

From the 

Data Display

 screen, press both

 

the

 Keypad

 

Left

 and 

Right

 buttons at the same time to 

display the first 

Function Menu

 screen. Next, use the 

Keypad

 buttons to move the arrow (>) until it 

points to the 

FAULTS

 function, and then press the 

ENTER

 button to display the 

Faults

 

Log

 screen.  

The 

Faults Log

 screen shows the monitor’s current fault status. If the 

fault is still present when the 

FAULTS

 function is selected, then the 

current cause of the fault is displayed along with the date and time it 
occurred. If the cause of the fault has been cleared, then the 

Faults Log

 

screen will show <0000> along with the date and time the fault was 

cleared. Use the 

Keypad

 buttons to scroll through the fault log. In the 

screens shown to the left, record #03 shows that a Sample Flow Fault (fault 

code <0800>) occurred on 07/11/05 at 08:17, while record #04 shows that 

the fault was cleared on 07/11/05 at 9:00.  

The cause of the fault is identified by a numeric fault code.  To convert the 

fault code into a text description of the fault, first press the 

ENTER

 button 

and then use the 

Keypad

 buttons to scroll through the display until the 

text description of the fault appears. If the fault code is a combination of 

two or more faults, then continue to use the 

Keypad

 buttons until all fault 

text descriptions have been displayed. For example, the fault code <1800> 

represents the combination of both a Sample Flow <0800> and a Purge 
Flow <1000> fault as shown to the left. 
The Fault Log can be cleared as described under the heading 

Clearing the 

PPM Log & Faults Log

 on Page 12. 

3.8.4 Fault Codes 

FAULT CODES ARE ADDITIVE

.  For example: 

A fault code of <0003> indicates that both a Box 

Temperature Fault <0001> and a Bench Temper-

ature Fault <0002> have occurred.  

<0001> Box Temperature Fault:

 Enclosure 

temperature is outside normal range (or IR detector 

has failed). Check that the monitor is not being 
subjected to extreme temperatures. Use the 

DIAG

 

function to check the Box Temperature.

 

<0002> Bench Temperature Fault:

 Optical 

bench is outside normal operating range (or IR 

detector has failed). Check that the monitor is not 
being subjected to extreme temperatures.  

<0004> Manifold Pressure Fault:

 The manifold 

pressure is outside its normal operating range (or 

IR detector has failed). Enter the 

DIAG

 

function and 

record ALL data. Call the factory with this 
information for further instructions. 

<0040> Fill Flow Fault:

 The purge-air bag’s 

pressure drop is outside expected limits. Check for 

a punctured bag or disconnected tubing. 

<0080> Over Range Fault:

 Monitor exposed to a 

gas level that exceeded 65,000 ppm.

 

<0100> Zero Filter Fault:

 The air-purge bag is 

contaminated with gas. Take monitor to a clean-air 
area and use the 

EMTYBAG

 and 

FILLBAG

 functions 

to decontaminate the air-purge bag.

 

<0200> Gain Set Fault:

 The digipot autotune 

sequence has failed. This fault will only occur on 

first boot up or after a firmware upgrade. Call the 

factory for further instructions. 

<0400> A/D Fault:

 A fault has occurred in the 

analog-to-digital circuitry. Contact the factory with 

this information for further instructions. 

<0800> Sample Flow Fault:

 Check for: 

A restriction in the gas-sample inlet or exhaust; a 
blocked internal filter; or a failed pump. 

<1000> Purge Flow Fault:

 Check for: 

A restriction in the gas-sample exhaust; a blocked 
internal filter; or a failed pump. Once the purge air 

stream has been restored, the monitor will return 

to normal operation after it completes a purge 
cycle. 

<2000> Bag Fill Fault:

 The purge-air bag did not 

fill within the expected time allotment. Check for a 

punctured bag or disconnected tubing. 

<4000> Zero Range Fault:

 The IR detector’s 

output voltage is out of tolerance. Enter the 

DIAG

 

function and record all data. Call the factory with 
this information for further instructions. 

<8000> Clipping Fault: 

The detector voltage may 

be out of tolerance. Use the 

DIAG

 

function to check 

the IR detector voltage. Call the factory with this 

information for further instructions. 

FAULT CODE<1800> 
12   SAMPLE FLOW 

FAULT CODE<1800> 
13    PURGE FLOW 

#03 <0800> @ 
07/11/05  08:17 

#04 <0000> @ 
07/11/05  09:00 

Содержание H-SEM

Страница 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Страница 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Страница 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Страница 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Страница 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Страница 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Страница 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Страница 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Страница 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Страница 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Страница 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Страница 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Страница 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Страница 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Страница 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Страница 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Страница 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Страница 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Страница 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Страница 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Страница 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Страница 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Страница 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Страница 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Отзывы: