background image

H-SEM 

Operation 

Instruction 3015-4584 

 

7

 

3.5 Display Screens 

3.5.1 Initial Power Up 

When the monitor is first powered up, all front panel lights turn ON and a splash screen appears showing 

the monitor’s current firmware revision level. After a brief moment the 

Warm Up 

screen is displayed 

along with the front panel 

ON 

light (green) blinking.  

 

 

The monitor requires 900 seconds (15 minutes) to warm up; after which, the 

ON

 light glows steady and 

the 

Data Display 

screen is displayed. 

3.5.2 Filling the Purge Air Bag 

WARNING:

 

Fill the purge-air bag in a clean, fresh-air environment that is at the same temperature 

and humidity as the area being checked for gas. DO NOT fill the bag in an area that is 

contaminated with refrigerant gas.

 

The following message is displayed when the monitor’s purge air bag is empty and needs refilling. Press 

the 

ENTER

 button to start the refill process. After the air bag has been filled, the monitor will 

automatically resume normal operation. 
 
 

3.5.3 Data Display Screen 

 
 
 

During normal operation, the 

Data Display

 screen shows when the monitor is performing the following 

two functions: 

MEASURE

 indicates that the monitor is actively measuring gas. In the screen shown above, 10 ppm of 

R-134A gas is currently being detected, and that a peak measurement of 20 ppm has occurred sometime in 

the past. The measurement cycle runs for 4 minutes. 

To reset the peak value to zero, press the 

ENTER

 button. 

PURGE

 is displayed when the monitor is resetting its infrared detector to a baseline of 0 ppm using the 

air stored in the internal purge-air bag. The purge cycle runs for 10 seconds. The monitor will operate for 

approximately 40 minutes before the purge-air bag needs refilling. 

3.5.4 Function Screens 

The Function

 

screens are used to display stored data and to set up the monitor. 

 
 
 

 
 

BACHARACH 

VERSION x.xx 

WARM UP
       900

MEASURE  00020pk 
   10ppm R134A 

>FILLBAG EMTYBAG 
 ALRMLVL AUDALRM 

>PPM LOG LOG INT
 FAULTS  DIAG    

>CUST K1 CAL 
 CUST K2 CUST K3 

PURGE    00020pk
   10ppm R134A

>SILENCE  P-CHK  

GAS      CLOCK 

PURGE AIR EMPTY 
<ENTER> TO FILL

RECHARGING PURGE
AIR...

Содержание H-SEM

Страница 1: ...H SEM Halogen Safe Entry Monitor Instruction 3015 4584 Operation Maintenance Rev 0 November 2005 Product Leadership Training Service Reliability...

Страница 2: ...ACCEPTED IN LIEU OF I ANY AND ALL OTHER WARRANTIES EXPRESS OR IMPLIED INCLUDING WITHOUT LIMITATION THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE AND II ANY OBLIGATION...

Страница 3: ...4 2 Charging the Battery 6 3 5 Display Screens 7 3 5 1 Initial Power Up 7 3 5 2 Filling the Purge Air Bag 7 3 5 3 Data Display Screen 7 3 5 4 Function Screens 7 3 6 Standby Mode 10 3 7 Working with t...

Страница 4: ...tents H SEM ii Instruction 3015 4584 4 MAINTENANCE 15 4 1 Disassembly 15 4 2 Internal Particulate Hydrophobic Filter Replacement 16 5 PARTS AND SERVICE 17 5 1 Replacement Parts 17 5 2 Service Centers...

Страница 5: ...d associated with the use of this equipment It calls attention to a procedure practice or condition or the like which if not correctly performed or adhered to could result in personal injury or death...

Страница 6: ...ting due to changes in air pressure Deflate the bag by selecting the EMTYBAG function as described on Page 8 1 7 Unpacking and Initial Checks Your Halogen Safe Entry Monitor was carefully inspected at...

Страница 7: ...e current gas level and the peak gas level detected in that area on its front panel LCD The monitor retains a log of previous readings that can be easily accessed at a later time for analysis Front pa...

Страница 8: ...onds Temperature Drift 1 5 ppm per C between purge cycles System Noise Less than 40dB A at 10 ft 3 m Operating Temp 32 to 122 F 0 to 50 C Ambient Humidity 5 to 90 RH non condensing Altitude Limit 6 56...

Страница 9: ...h air that is of the same temperature and relative humidity as the area being checked for gas DO NOT fill the purge air bag in an area that is contaminated with refrigerant gas After the purge air bag...

Страница 10: ...le alarm will reactivate if the fault condition is not cleared within the time period set by the SILENCE function Page 9 Both the FAULT light and audible alarm will automatically turn OFF after the ca...

Страница 11: ...ill automatically resume normal operation 3 5 3 Data Display Screen During normal operation the Data Display screen shows when the monitor is performing the following two functions MEASURE indicates t...

Страница 12: ...g ON AUDALRM Allows the monitor s internal audible alarm to be associated with any function of the monitoring system Factory default is OFF Use the Keypad to select the desired audible alarm function...

Страница 13: ...pe as specified on the sales order or a custom gas can be programmed into the monitor at a later time using the CUST K1 K2 K3 function Use the Keypad to select the desired gas type and then press ENTE...

Страница 14: ...panel ALARM light will flash and the audible alarm if activated refer to AUDALRM function Page 8 will start beeping The ALARM light and audible alarm will automatically turn OFF once the detected gas...

Страница 15: ...eme temperatures Use the DIAG function to check the Box Temperature 0002 Bench Temperature Fault Optical bench is outside normal operating range or IR detector has failed Check that the monitor is not...

Страница 16: ...the First Diagnostic screen the user can toggle between displaying Bench Voltage Detector Temperature and Noise Box Temperature by pressing the Keypad Right button Bench Voltage This is the current pe...

Страница 17: ...e Data Display screen press both the Keypad Left and Right buttons at the same time to display the first Function Menu screen Next use the Keypad buttons to move the arrow until it points to the P CHK...

Страница 18: ...temperature drift IR source aging and pressure changes in order to keep the unit within factory accuracy specifications 3 13 1 Calibration Procedure The calibration factor is determined by sampling a...

Страница 19: ...l chassis as follows Items Required Medium Phillips head screwdriver Procedure 1 Remove monitor and its battery pack from the soft carrying case 2 Unplug battery pack from monitor 3 Remove a total of...

Страница 20: ...1 2 See illustration below and locate the internal particulate hydrophobic filter 3 Pull off tubing from both ends of filter and remove filter from instrument 4 Attach tubing to new filter being sure...

Страница 21: ...rvice 5 1 Replacement Parts Item Description Part Number Halogen Safe Entry Monitor 3015 4484 External Filter 3015 3420 Internal Particulate Filter 07 1654 Battery Pack Charger 3015 4654 Battery Charg...

Страница 22: ...il help bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xic...

Страница 23: ...H SEM Parts and Service Instruction 3015 4584 19 Notes...

Страница 24: ...1 Hunt Valley Circle New Kensington PA 15068 PH 724 334 5000 Toll Free 800 736 4666 FAX 724 334 5001 Website www bacharach inc com E mail help bacharach inc com Printed in U S A Registered Trademark o...

Отзывы: