background image

Instruction 24-9428

2

Fyrite Tech

1.2  Operational Overview

Pressng the 

I/O

 button turns the analyzer ON. Note that there s a 60 sec-

ond warm-up perod that must elapse before the analyzer can be used.

After warm-up s complete, choose the fuel code that corresponds to the 

fuel beng burned by the applance beng tested. Then press the 

ENTER 

button to place the analyzer nto ts Run Mode and begn the combuston 

testng process.

Begin testing by inserting the analyzer’s probe tube into the flue-gas 

stream of the applance under test. Each analyzer model wll contnuously 

monitor the flue gas and display measured and calculated values that 

are relatve to the combuston process. These values are dsplayed on the 

analyzer’s LCD, and are chosen for dsplay by pressng the Increment 

(

)

 

and Decrement 

(

)

 buttons.

The Fyrte Tech 60 has the capablty of sendng the current combuston 

test to an optonal prnter by pressng the 

HOLD

 button twce.

Pressng the 

HOLD 

button durng a test wll freeze all measured and 

calculated values at ther current values. Pressng the 

ENTER 

button 

resumes testng.

A backlght enables an operator to read the dsplay n dmly-lt areas. 

Pressng the 

LIGHT 

button turns the backlght ON and OFF. The backlght 

wll automatcally turn OFF after 10 mnutes of keyboard nactvty.

A power-saver functon wll shut the analyzer OFF after 20 mnutes of 

keyboard nactvty. Note, however, that the power-saver functon s ds-

abled f the detected O

2

 value s less than 17.9% or the CO value s more 

than 50 ppm.

Press the 

I/O

 button to turn the analyzer OFF. Note that there s a 5 sec-

ond delay before the analyzer actually turns OFF, durng whch tme the 

unt can be kept ON by pressng the 

ENTER 

button. 

If the analyzer s turned OFF whle a hgh level of CO s stll present 

wthn the unt, the pump remans runnng and the unt wll not turn OFF 

untl the detected CO level drops below 50 ppm.

Содержание Fyrite Tech 50

Страница 1: ...Fyrite Tech Model 50 60 Combustion Gas Analyzer Instruction 24 9428 Operation Maintenance Rev 4 August 2005 Product Leadership Training Service Reliability...

Страница 2: ...the Buyer shall be limited to those provided herein to the exclusion of any and all other remedies includ ing without limita tion incidental or consequen tial damages No agree ment varying or extendi...

Страница 3: ...Date Setup Model 60 only 11 3 5 6 Year Setup Model 60 only 11 3 5 7 Selecting Printer Protocol Model 60 only 12 3 5 8 Exiting Setup 12 4 0 OPERATION 13 4 1 Performing a Test 13 4 1 1 Analyzer Turn On...

Страница 4: ...CO2 Excess Air Screen 25 5 0 CALIBRATION MAINTENANCE 27 5 1 Entering the Calibration Mode and Testing the Display Segments 27 5 2 Ambient Temperature Calibration 28 5 3 Stack Thermocouple Channel Cali...

Страница 5: ...Temperature Ambient Temperature Carbon Monoxide Calculations Combustion Efficiency Excess Air Carbon Dioxide CO air free Print Capability X X X X X X X X X X X X X X X The printer used with a Fyrite...

Страница 6: ...Fyrite Tech Instruction 24 9428 iv Notes...

Страница 7: ...l printer via an infrared port using either HP or IrDA protocol Fyrite Tech 50 Fyrite Tech 60 Measures Measures Oxygen O2 Oxygen O2 Stack temperature Stack temperature Ambient temperature Ambient temp...

Страница 8: ...e current combustion test to an optional printer by pressing the HOLD button twice Pressing the HOLD button during a test will freeze all measured and calculated values at their current values Pressin...

Страница 9: ...fficiency 0 1 to 99 9 Flue Gas CO2 content 0 0 to a fuel dependent maximum value in Flue Gas CO Air Free content Model 60 0 to 9 999 ppm Excess Air 0 to 400 Fuel Selection F1 thru F4 denotes the fuel...

Страница 10: ...m whichever is greater Stack Temperature 4 F between 32 255 F 2 C between 0 124 C 6 F between 256 480 F 3 C between 125 249 C 8 F between 481 752 F 4 C between 250 400 C Ambient Temperature 4 F betwee...

Страница 11: ...ent charge prior to each use Replace the batteries if the low battery symbol appears in the lower right corner of the screen To replace the bat teries 1 Remove battery cover from back of analyzer 2 Re...

Страница 12: ...itting 2 Push the flue gas thermocouple into the T STACK jack Important DO NOT force thermocouple connector The con nection tabs are different sizes allowing the connector to fit in only one way Inspe...

Страница 13: ...Instruction 24 9428 Fyrite Tech Figure 3 2 Connecting the Probe T STACK GAS THERMOCOUPLE PROBE STOP Optional WATER TRAP FILTER ASSEMBLY...

Страница 14: ...elect Screen when held down for 2 seconds while in the Run Mode Sets up the analyzer to be placed into its Calibration Mode when held down with the analyzer OFF Used in conjunc tion with the I O butto...

Страница 15: ...ing Setup 1 With the analyzer turned OFF press and hold down the HOLD button 2 Press the I O button 3 Release both buttons 4 The analyzer is now in its Setup Mode Refer to Sections 3 5 2 thru 3 5 7 fo...

Страница 16: ...ent condi tions during start up A2 Auto Zero With the CO channel set to auto zero the CO sensor is zeroed to the ambient CO level during start up Important When using this mode the analyzer must be tu...

Страница 17: ...ot already done enter Setup per Section 3 5 1 Repeatedly press ENTER until the first Date Setup Screen is displayed 2 Press the or button until the correct month is displayed 3 Press ENTER to change t...

Страница 18: ...ction 3 5 1 Repeatedly press ENTER until one of the following screens is displayed 2 Press the or button to select the desired printer protocol 3 Press ENTER to move to next Setup Screen or press I O...

Страница 19: ...robe is properly positioned turn ON the ana lyzer by pressing its I O button 3 Wait for the analyzer to complete its 60 second warm up period after which the Fuel Screen is displayed If errors were de...

Страница 20: ...tip directly into the top of the fire tube below the diverter 80 Efficiency Fan Assist or Power Vented Combustion testing of fan assist or power vented furnaces boilers should be done through a hole...

Страница 21: ...Instruction 24 9428 15 Fyrite Tech 90 Eff Condensing Furnace Boiler Atmospheric or Gravity Vented Boiler 80 Eff Fan Assist or Power Vented Furnace Boiler...

Страница 22: ...es all readings and stops the pump The and buttons can now be used to scroll through the dis plays and view all test values at the moment the HOLD button was pressed Pressing ENTER restarts the pump a...

Страница 23: ...Instruction 24 9428 17 Fyrite Tech Figure 4 1 Printer Alignment Sample Printout BACH AR AC H I NC F YR IT E TE CH A NA LY ZE R...

Страница 24: ...r the analyzer has been purged with fresh air turn it OFF by pressing the I O button The unit will count down from 5 before turning OFF Press ing the ENTER button however will stop the count down and...

Страница 25: ...yzer s internal components to become soaked in condensate for long periods of time Before each use inspect the filter element of the water trap filter as sembly Replace the filter if it looks dirty Wh...

Страница 26: ...e Probe A rigid stainless steel probe with handle connected to a flexible hose with integral water trap filter can be used to draw a gas sample into the ana lyzer from the room grilles diffusers and f...

Страница 27: ...Tech 50 Fuel Select F1 Natural Gas F2 Oil 2 F3 LPG F4 Kerosene O2 Ambient Air Temperature Stack Temperature Efficiency CO2 Excess Air Fuel Select F1 Natural Gas F2 Oil 2 F3 LPG F4 Kerosene CO O2 CO Ai...

Страница 28: ...ayed after warm up or can be accessed from any Run Mode Screen by holding down the ENTER button for 2 seconds The Fuel Screen is where an operator chooses which fuel is to be used by the analyzer to c...

Страница 29: ...cond delay 4 8 4 CO O2 Screen Model 60 only The CO display is the measured ppm amount of carbon monoxide present in the flue gas stream Note however that the amount of CO displayed may be diluted by s...

Страница 30: ...h a 5 second delay 4 8 6 Ambient Air Temperature Screen The AMBIENT reading is the room temperature in either F or C as mea sured by a temperature sensor located inside the analyzer Note The combustio...

Страница 31: ...eeze LIGHT Toggles backlight ON OFF I O Turns analyzer OFF with a 5 second delay 4 8 8 CO2 Excess Air Screen The CO2 reading is a calculation of the percentage of carbon dioxide that is being produced...

Страница 32: ...etected carbon mon oxide level is above 50 ppm during the warm up cycle Note that if the CO channel is set up for manual calibration the analyzer does not auto zero the CO sensor during warm up and th...

Страница 33: ...e analyzer turned OFF place the unit in fresh ambient air 2 Press and hold down the ENTER button and then press and release the I O button Observe that all LCD segments are turned ON 3 Release the ENT...

Страница 34: ...ibration screen or hold down ENTER for 2 seconds to store the displayed value and enter the Run Mode or press the I O button to exit the Calibration Mode and turn OFF the analyzer without sav ing the...

Страница 35: ...l Required Cylinder of 100 Nitrogen P N 9550 0049 Calibration Kit P N 24 7059 Procedure 1 With the analyzer sampling fresh air enter the Calibration Mode as described in Section 5 1 Then repeatedly pr...

Страница 36: ...t P N 24 7059 If the analyzer will be primarily used for flue gas testing we suggest calibrating your Fyrite Tech 60 with 500 ppm CO calibration gas If the analyzer will be primarily used for ambient...

Страница 37: ...isplayed value to match the CO concentration stamped on the gas cylinder 8 Press ENTER to store the new calibration values and move to the next calibration screen or hold down ENTER for 2 seconds to s...

Страница 38: ...lter Element when dirty Material Required Filter Element P N 07 1644 Small Flat Blade Screwdriver Procedure 1 Pull apart the Water Trap Filter assembly see Figure 5 2 2 Pry apart the Filter Chamber Th...

Страница 39: ...electrical connector from the circuit board Then push down twist counterclockwise and then pull the O2 sen sor out of its socket see Figure 5 3 Tip To obtain a better grip on the O2 sensor it may be n...

Страница 40: ...laying it face down to the left of the body d Remove the battery connector from top of circuit board Then carefully remove the circuit board laying it face down in the top cover 2 Gently pull CO sens...

Страница 41: ...move the circuit board laying it face down in the top cover 2 Remove the pump motor s electrical connector from the circuit board 3 Unscrew the pump s hold down clamp and remove it from the pump see F...

Страница 42: ...Instruction 24 9428 36 Fyrite Tech Figure 5 3 Back Case Component Locations CO SENSOR GASKET OXYGEN SENSOR SOCKET PUMP HOLD DOWN CLAMP IN IN OUT OUT...

Страница 43: ...Instruction 24 9428 37 Fyrite Tech Figure 5 4 CO Sensor Location CO SENSOR To BATTERY To OXYGEN SENSOR T STACK To PUMP RED WIRE RED WIRE RED WIRE...

Страница 44: ...pack of 3 07 1644 6 2 Accessories STANDARD ACCESSORIES Battery AA Alkaline 204 0004 Instruction Manual 24 9428 Soft Carrying Case 24 1267 Quick Start Guide 24 9429 OPTIONAL ACCESSORIES Calibration Ki...

Страница 45: ...p bacharach inc com Canada Bacharach of Canada Inc 250 Shields Court Unit 3 Markham Ontario L3R 9W7 Canada Phone 905 470 8985 Fax 905 470 8963 Email bachcan idirect com M xico Bacharach de M xico Play...

Страница 46: ...Instruction 24 9428 40 Fyrite Tech Notes...

Страница 47: ...Instruction 24 9428 41 Fyrite Tech Notes...

Страница 48: ...Headquarters 621 Hunt Valley Circle New Kensington PA 15068 Ph 724 334 5000 Fax 724 334 5001 Toll Free 800 736 4666 Website www bacharach inc com E mail help bacharach inc com Printed in U S A...

Отзывы: