Instruction 24-9428
31
Fyrite Tech
5. Set up the Calbraton Kt wth 100 or 500 ppm CO as descrbed n
the nstructons suppled wth the kt.
6. Connect the tubng of the Calbraton Kt to the nlet of the an-
alyzer. Then adjust the regulator for approximately 2 SCFH of
excess flow (see Figure 5.1).
7. After the analyzer has stablzed (2 to 3 mnutes), use the
and
buttons to set the dsplayed value to match the CO concentraton
stamped on the gas cylnder.
8. Press
ENTER
to store the new calbraton values and move to the
next calbraton screen, or hold down
ENTER
for 2 seconds to store
the new calbraton values and enter the Run Mode, or press the
I/O
button to ext the Calbraton Mode and turn OFF the analyzer
wthout savng the changes.
9. Turn off gas flow and disconnect tubing from analyzer.
Figure 5-1. Calibration Kit Hookup
LEAVE TOP
PORT OPEN
Parts Shown:
1. Gas Cylinder
2. Regulator*
3. Tubing*
4. Tee*
5. Flowmeter*
6. Fitting, Gas
* Contained in Calibration Kit