Instruction 24-9395
3
Fyrite Pro
2.0 TeChNICAl ChARACTeRISTICS
The Analyzer Directly Measures and Displays:
• Ambent Temperature ............................32 to 212 °F (0 to 100 °C)
• Flue Gas O
2
content ................................0.0 to 20.9% O
2
• Flue Gas Temperature ...........................0 to 999 °F (–18 to 537 °C)
• Flue Gas CO content ..............................0 to 2,000 ppm CO
• Draft ........................................................–9.99 to 27 nWC
(–9.99 to 67 mb/HPa)
The Analyzer Computes and Displays:
(when the measured oxygen level s below 17.9%)
• Combuston Effcency ............................0.1 to 99.9%
• Flue Gas CO
2
content .............................0.1 to a fuel dependent
maxmum value n %
• Flue Gas CO Ar-Free content ...............0 to 9,999 ppm
• Excess Ar ...............................................0 to 400%
Fuel Selection:
(F1 thru F4 denotes the fuel selected as dsplayed on the LCD)
• Natural Gas (F1)
• Ol #2 (F2)
• LPG (F3)
• Kerosene (F4)
Normal Operating Conditions:
Temperature:
• Analyzer .................................................32 to 104 °F (0 to 40 °C)
• Probe .......................................................1,000 °F max (538 °C) at
5" nserton
Humdty .......................................................15 to 90% RH,
Non-Condensng
Power Requrements ....................................Four dsposable ‘AA’ Alkalne
batteres (an optonal AC adapt-
er s avalable that elmnates
the need for batteres)
Содержание 100
Страница 6: ...Fyrite Pro Instruction 24 9395 iv Notes...
Страница 22: ...Instruction 24 9395 16 Fyrite Pro Notes...
Страница 46: ...Instruction 24 9395 40 Fyrite Pro Notes...
Страница 58: ...Instruction 24 9395 52 Fyrite Pro Figure 5 4 Case Components Plumbing Schematic...
Страница 60: ...Instruction 24 9395 54 Fyrite Pro Notes...