Instruction 24-9395
46
Fyrite Pro
6. Connect the tubng of the Calbraton Kt to the GAS nlet of the an-
alyzer. Note that the gas fttng (Item 6) s only requred on Fyrte
Pro Models 105 and 125.
7. Adjust the regulator for approxmately 2 SCFH of excess flow (see
Fgure 5.1).
8. After the analyzer has stablzed (2 to 3 mnutes), use the
and
buttons to set the dsplayed value to match the CO concentraton
stamped on the gas cylnder.
9. Press
eNTeR
to store the new calbraton values and move to the
next calbraton screen, or hold down
eNTeR
for 2 seconds to store
the new calbraton values and enter the Run Mode, or press the
I/O
button to ext the Calbraton Mode and turn OFF the analyzer
wthout savng the changes.
10. Dsconnect tubng from analyzer and turn off gas flow.
Figure 5-1. Calibration Kit Hookup