Specifications
104
DC Current
Rate
Range
Resolution
Full Scale
Reading
Accuracy
(1 year)
Burden Voltage
(1)
& Shunt Resistor
Slow
12.0000mA
120.000mA
1.20000 A
12.0000 A
(3)
0.1 μA
1 μA
10μA
100μA
11.9999
119.999
1.19999
11.9999
0.05+0.008
(2)
0.05+0.004
(2)
0.10+0.004
0.25+0.004
<0.15 V / 10.1
Ω
<1.5 V / 10.1
Ω
<0.3 V / 0.1
Ω
<0.15 V / 10 m
Ω
Med
12.0000mA
120.000mA
1.20000 A
12.0000 A
(3)
0.1 μA
1 μA
10μA
100μA
11.9999
119.999
1.19999
11.9999
0.05+0.015
(2)
0.05+0.008
(2)
0.10+0.008
0.25+0.008
<0.15 V / 10.1
Ω
<1.5 V / 10.1
Ω
<0.3 V / 0.1
Ω
<0.15 V / 10 m
Ω
Fast
12.000mA
120.00mA
1.2000 A
12.000 A
(3)
1 μA
10μA
100μA
1 mA
11.999
119.99
1.1999
11.999
0.10+0.02
(2)
0.10+0.02
(2)
0.10+0.02
0.25+0.02
<0.15 V / 10.1
Ω
<1.5 V / 10.1
Ω
<0.3 V / 0.1
Ω
<0.15 V / 10 m
Ω
(1)
Typical voltage across the input terminals at full scale reading.
(2)
Use REL function
(3)
In 12A range,>10-12 A is readable for 20 seconds maximum
TRUE RMS AC CHARACTERISTICS
AC Voltage
Resolution and Full Scale Reading
Range
Resolution
Full Scale Reading
Slow & Med
Fast
Slow & Med
Fast
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
1 μV
10 μV
100μV
1 mV
10 mV
10 μV
100
μV
1 mV
10 mV
100 mV
119,999
1.19999
11.9999
119.999
757.50
(1)
119.99
1.1999
11.999
119.99
757.5
(1)
(1)
1% over- range (757.5V) is readable at 750 V range.
Accuracy
Rate
Range
Accuracy (1 year) ± (% of r % of range), 23°C ±5°C
(1)
10~20 Hz
20~50 Hz
50~20 kHz
20~50 kHz
50~100 kHz
Slow
120.000 mV
1.20000 V
12.0000 V
120.000 V
750.00 V
1.50+0.100
1.50+0.100
1.50+0.100
1.50+0.100
1.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.50+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.10+0.100
0.30+0.150
0.30+0.100
0.30+0.100
0.30+0.100
0.30+0.100
(2)
1.0+0.150
1.0+0.100
1.0+0.100
1.0+0.100
1.0+0.100
(1)