Specifications
107
Maximum Input Protection
: 1000VDC or 750VAC for all ranges.
Open circuit voltage
: Maximum voltage is 13.3 VDC for 120
Ω,
1.2
kΩ,
12
Ω,
12
MΩ
and 120
MΩ
ranges.
Maximum voltage is 7 VDC for 120
kΩ and 1
.2M ranges.
Continuity
Accuracy ± (% of r % of range), 23°C ±5°C
Rate
Range
Resolution
Full Scale
Reading
Test Current
Accuracy
(1 year)
Fast
1 kΩ
100
mΩ
999.9
1 mA
0.10+0.020
Maximum Input Protection: 1000VDC or 750VAC for all ranges.
Open circuit voltage
: <13.3 VDC
Test Current
: Approximately 1 mA DC
Threshold resistance
: From 1
Ω
to 1000
Ω, default of power on is 10
Ω
Diode
Accuracy ± (% of r % of range), 23°C ±5°C
Rate
Range
Resolution
Full Scale
Reading
Test Current
Accuracy
(1 year)
Med
3.0000 V
10.0000 V
10.0000 V
100
μV
100
μV
100
μV
2.9999
10.0000
10.0000
1 mA
100
μA
10
μA
0.030+0.020
0.030+0.020
0.030+0.020
Maximum Input Protection: 1000VDC or 750VAC for all ranges.