13
Cleaning / Battery
Regular checking, cleaning,
disinfection
Check regularly
Self-check, audible alarm, process- and
alarm control. Use only according to
Instructions for Use.
Cleanliness, completeness, damage.
Function of staff call (if connected).
Cleaning
With standard household cleaners, soap and
water solution.
Wet disinfection
Only use surface disinfection agents. Do not
use spray disinfectant at mains connection.
Before operation the device allows to vent for
at least 1 min. B. Braun recommends
Meliseptol.
Service
Servicing must be carried out exclusively by
personnel instructed by B. Braun.
Carry out safety check every 24 months.
Use only original spare parts and
accessories.
12 V mains operation or
battery
Check mains voltage as per type plate.
Plug in power lead at rear (screw in 12 V
lead in ambulance car).
In case of power failure or if 12 V DC/230
V AC is not connected the unit automatically
will be switched to integrated rechargeable
battery.
Charge Battery
Charge battery in case of
-
first use,
-
battery alarm,
-
non-use for more than 2 months.
Battery is charged - even during infusion -
if 12 V or 230 V AC are connected.
Charging time
Approx. 16 hours. Longer charging is not
detrimental.
Capacity
A fully charged battery is sufficient for more
than 3.5 hours with delivery rates < 25 ml/h.
Rechargeable batteries ageing
After 2 years, the original capacity is only
approx. 50 %.
The life time of the battery can be exten-
ded by completely discharging from time to
time and recharging afterwards.
Inspection on Delivery
Despite careful packaging, the risk of
transport damage cannot be entirely
prevented. Upon delivery, please check that
nothing is missing. Do not use a damaged
device. Contact the service department.
Items included
Perfusor fm, mains lead, Pole clamp,
Instructions for Use.