12
Reasons for alarm
When display indicates "device alarm" an
audible alarm sounds permanently.
Press ON/OFF-key repeatedly until display
indication "do not press any key until display is
off". Pump switches off automatically after a
few seconds.
Switch on device again. In case of repeated
device alarm inform service.
Remedy failure and restart infusion
Pressure alarm
Occlusion? - Check the free flow and insert
infusion line free of kinks. Bolus is automatically
reduced.
Press START for renewed infusion.
Battery alarm
+ -
Battery pre-alarm 3 min. before battery is
discharged.
Then battery alarm:
- Yellow battery indicator flashes.
- Red alarm indicator is on.
- Audible alarm every 4 s.
- Operating indicator off.
- Staff call.
Switch off unit. Connect to mains.
Syringe alarm
Fault on syringe replacement? Manipulation of
syringe holder during infusion?
Pull syringe holder.
Syringe pre-alarm
3* min before syringe empty:
- Display: Syringe pre-alarm.
- Audible alarm every 9 s.
* Time can be set by our Service staff to between 3 and 30 min.
If syringe empty:
- Display: Syringe empty.
- Operating indicator off.
- Audible alarm every 4 s.
- Staff call.
Standby alarm
Alarm after pre-selected pause?
To terminate pause: Press key beneath END.
To extend pause: Press key beneath ON.
Further alarms/displays
Display: No rate? Set rate.
Operating alarms
Device Alarms