Change Syringe Concentration
with Dosage Calculation Activated
Note: Always out of basic menu.
If necessary, press END until display basic menu
is VOL TIME -- SF.
Press key below SF repeatedly until DOSAGE
CALC. ON appears and confirm
with OK.
Clear syringe concentration with C.
Enter new value and confirm.
Change Dosage from Basic Menu
Press key below SF repeatedly until DOSAGE
CALC. ON appears and confirm
with OK.
Confirm concentration with OK.
If active, confirm weight with OK.
Change values (move flashing star as
described) and confirm entries with OK.
Check entered or calculated rate for
plausibility and confirm with key below RATE.
Rate appears in red display.
Start Infusion.
Switch off dosage calculation
The dosage calculation remains activated
until it is switched off in Special Functions. If
the Perfusor fm is switched off in the mean-
time all previous values except the body weight
are maintained.
Switch off from Basic Menu
Press key below SF repeatedly until display
DOSISCALC.ON.
Press DOSISCAL.OFF.
Dosage calculation is deactivated.
With key below END back to basic menu.
Interface
Interface Operation:
Connection to interface input via MFC-plug.
Interface descriptions available from B. Braun.
Remote Control
Via fm controller possible. When using a
commercially available external computer this
must satisfy the requirements acc. to IEC/EN
60601-1 as well as the single-fault fail-safe
condition acc. to IEC/EN 60513.
Send Proposal
The delivery rate and a drug can be entered in
the Perfusor fm as "proposal" via an external
computer. Both items of data must be checked
on the Perfusor fm and acknowledged.
Documentation
All operating data of the Perfusor fm can be
requested and logged via external computer.