5

Potential equalisation

Connect with CF-mode.

CF-application (cardiac floating),

defibrillator-proof

Pole clamp, back
Attach to infusion stand. Fit
Perfusor fm from above and
engage. To release, press
black button.

Delivery rate

Keypad for input

Correction/CLEAR key
Resets display to 000.

0

Main switch ON/OFF
Press for 2 seconds to switch off.

Remaining battery capacity
With device switched off and power lead
disconnected: Briefly press 

 .

Battery capacity (and ward if envisaged)
is indicated after 3 seconds.

Suppress alarm tone for 2 min

START/STOP
Start/interruption of infusion, e.g. for altering
delivery rate.

Aluminium casing
Easy upkeep, drip-proof and resistant to
disinfectant.

Memory-card window
Scope of functions can be extended for
exemple by way of a PCA memory card (please
ask for information).

Handle for easy transport

Multi-function-connector (MFC)*
Connecting for staff call, ambulances (12 V DC)
and interface for fm anaesthesia/intensive.

Mains connection
Connection for power lead.Automatic
switching to battery in the event of power
failure. Mains fuses directly above mains
connection.

Power supply
For operation with fluid manager system.

Infrared interface
For operation with fluid manager system.

Slide-in memory card unit

Alarm Indicator

Lights in alarm situation.

Operating Indicator

Indicates delivery.

Battery Check

Lights when using battery. Flashes with battery
pre-alarm.

Mains Check

Lights during mains operation. Battery is
charged.

* As of unit no. 20001. Previously separate connections

for staff call and 12 V supply.

Содержание Perfusor fm

Страница 1: ...1 Perfusor fm 0123 Software PFAE Instructions for Use...

Страница 2: ...stallation in medically used rooms must comply with the appropriate regulations e g VDE 0100 VDE 0107 or IEC publications Observe national specifications and deviations The infusion lines should be pl...

Страница 3: ...roducible infusion of medication in large and small quantities The device permits extremely exacting delivery rate settings and thus optimal adaptation to any given requirements The drive system ensur...

Страница 4: ...nge in position To remove syringe Pull and move to the left SF Special functions If released successive call up of Dosage Calculation Bolus function Standby function Drug selection Occlusion pressure...

Страница 5: ...disinfectant Memory card window Scope of functions can be extended for exemple by way of a PCA memory card please ask for information Handle for easy transport Multi function connector MFC Connecting...

Страница 6: ...atient Set delivery rate Values between 0 1 and 200 ml h Check display Correction Press C enter rate again Choose additional settings as described below For automatic rate calculation Do not enter rat...

Страница 7: ...d as ml or a new target volume has to be set Target Time Time Pre selection Press key beneath TIME Enter target time via keypad and confirm with TIME 50 min 5 0 2 h 30 min 2 3 0 After confirming inste...

Страница 8: ...ues are displayed Standby Pause In case of extended interruptions set values are maintained Press STOP Press key beneath SF until Standby is displayed Press key beneath ON Select special functions SF...

Страница 9: ...han 2 hours is displayed after 16 hours charging time Drug Display 9 selectable drug names can be stored input via service program only Select with key beneath Drug is displayed also during infusion K...

Страница 10: ...numbers with OK Entry of syringe concentration per 1 ml or per syringe filling For dosage by body weight enter body weight and confirm Otherwise confirm 0 kg With OFF complete termination of all dosag...

Страница 11: ...t is switched off in Special Functions If the Perfusor fm is switched off in the mean time all previous values except the body weight are maintained Switch off from Basic Menu Press key below SF repea...

Страница 12: ...s discharged Then battery alarm Yellow battery indicator flashes Red alarm indicator is on Audible alarm every 4 s Operating indicator off Staff call Switch off unit Connect to mains Syringe alarm Fau...

Страница 13: ...ead in ambulance car In case of power failure or if 12 V DC 230 V AC is not connected the unit automatically will be switched to integrated rechargeable battery Charge Battery Charge battery in case o...

Страница 14: ...made of syringes disposables other than those stated in the order data 10 ml Omnifix Delivery rate 1 ml h 10 ml Omnifix 5 ml h flow Pressure flow flow deviation 20 ml OPS 1 ml h deviation Max bolus v...

Страница 15: ...and type of syringe 20 ml OPS Delivery rate 10 ml h flow 50 ml OPS Delivery 1 ml h flow 50 ml OPS Delivery rate 25 ml h flow 20 ml OPS 10 ml h deviation 50 ml OPS 1 ml h deviation 50 ml OPS 25 ml h de...

Страница 16: ...0 4 A slow blow IEC 127 External extra low voltage 12 V Staff call Max 24 V 1 A 24 VA Arbitrary connection polarity VDE 0834 Interference suppression EMC EN 55011 CISPR 11 EN 60601 1 2 IEC 601 1 2 Ti...

Страница 17: ...Manufacturer Display Type Syringe size Reference article number 10 12 ml 20 ml 50 60 ml B Braun Perfusor 8728615 8728810 Proinjekt 8728917 Omnifix 4617100 4617207 4617509 Euroject 4630505 Becton B D P...

Страница 18: ...st achieved by calling in Service Individual consultation enables allowance to be made for the specific conditions in a given hospital A complete service manual can be provided if required however onl...

Страница 19: ...Luer lock connectors pressure resistant 75 cm 872 2870 M PVC with loose lock nut on patient end pressure resistant 150 cm 872 2994 PE PE with Luer lock connectors pressure resistant 150 cm 872 2935 S...

Страница 20: ...rawing no M 637 010 014 F04d Raw material no 3891009 8 05 01 Printed on pulp bleached 100 chlorine free B Braun Melsungen AG Postfach 1120 D 34209 Melsungen Tel 05661 71 0 Fax 05661 71 20 44 HOSPITAL...

Отзывы: