3
Safe operation
The MINOP TREND handle can be combined with the
following suction/irrigation sheaths:
•
MINOP TREND suction/irrigation sheath 0° and
MINOP TREND endoscope 0°
•
MINOP TREND suction/irrigation sheath 30° and
MINOP TREND endoscope 30°
When irrigation button
4
is pressed, the field of view
of the adapted MINOP TREND endoscope
5
is rinsed
and thus freed from any blood and secretion.
¾
To rinse MINOP TREND endoscope
5
:
Press irrigation button
4
down to the stop.
Suction is interrupted while irrigation is active.
¾
To stop rinsing MINOP TREND endoscope
5
:
Release irrigation button
4
.
The valve automatically switches to continuous
suction and interrupts the irrigation.
Note
Suction is permanently active, except for the short
periods when irrigation is activated.
Disassembling
Dismounting the MINOP TREND suction/irrigation
sheath from the MINOP TREND handle
¾
Turn MINOP TREND suction/irrigation sheath
2
with MINOP TREND endoscope
5
inserted while
pulling it out rearwards from MINOP TREND
handle
3
.
Dismounting the tubes and the irrigation button
Fig. 1
¾
Remove the tubes from the tube connectors.
¾
Unscrew irrigation button
4
.
Dismounting adapter for holding arm RT099R
¾
Undo the cap nut.
¾
Remove adapter for holding arm RT099R.
Removing the MINOP TREND endoscope from the
MINOP TREND suction/irrigation sheath
¾
Unscrew the light cable from MINOP TREND
endoscope
5
.
¾
Remove the camera from the eyepiece funnel of
MINOP TREND endoscope
5
.
¾
Unlock the clamping lever of MINOP TREND suc-
tion/irrigation sheath
2
and remove MINOP TREND
endoscope
5
.