9
Stage
III
¾
Fully immerse the product in the disinfecting
solution. Make certain that all accessible surfaces
are moistened.
¾
Clean the product with a suitable cleaning brush
(e.g. PM995200) until all visible residues have been
removed from the surface.
¾
Brush through all surfaces that are not accessible
to visual inspection, e.g. in products with hidden
crevices, lumens or complex geometries, for at
least 5 min or until no more residues can be
removed. Mobilize non-rigid components, e.g. set
screws, links, etc. during cleaning.
¾
After that, thoroughly rinse (at least 5 times) with
the cleaning solution these difficult to access areas
of the product, using a single-use syringe (20 ml).
¾
To avoid the risk of corrosion, do not use metal
brushes or other abrasives for cleaning the product.
Stage
IV
¾
Rinse the product completely (all accessible
surfaces) under running water.
Stage
V
¾
Dry the product with compressed air.
Stage
VI
¾
Rinse the product completely (all accessible
surfaces) under running water.
¾
Allow water to drip off for a sufficient length of
time.
Stage
VII
¾
Dry the product completely with compressed air.
Содержание Aesculap Neurosurgery
Страница 2: ......
Страница 3: ...4 3 5 2 1 6 5 2 6 3 4 7 ...