6-6
6.2 Input of simulated signals from calibrator
Introduction
A special calibrator has been designed for the electromagnetic flowmeter. This special
calibrator can generate the same kind of signal as the flow rate signal produced by
the flowtube. This simulated signal can be used for checking relevant functions of the
transmitter.
Getting started
Prepare the following devices and components.
ȲA special calibrator (MGZ13) and a special cable
ȲA digital multimeter
Ȳ250-Ω resistance
Checking method
Check the relevant functions in accordance with F1X 1000 Flow Meter Calibrator User's
Manual
(CM2-F1X100-2001).
Содержание MagneW Plus+ MGG14C
Страница 6: ...iv ...
Страница 8: ...vi ...
Страница 11: ...1 3 Figure1 3 System using a Foundation Fieldbus ...
Страница 14: ...1 6 The following table is a summary of the functions of each of the keys ...
Страница 20: ...1 12 CALIBRATION OFF SHIPPING INFO OFF MAINTENANCE MODE MODE RETURN BASIC SETUP Figure1 10 MAINTENANCE MODE ...
Страница 26: ...2 6 ...
Страница 42: ...4 6 ...
Страница 78: ...A 10 ...
Страница 105: ......
Страница 108: ......
Страница 110: ......