6-9
6.4 Gain adjustment
Measuring method using LUI
Step
Procedure
Screen
1
Enter CALIBRATION MODE and display the screen shown
on the right.
0.0000
2
Set the cursor at the position of READY and
press the
key to start gain calibration at the
zero point (0 m/s).
When the screen shown on the right is displayed, the gain
calibration at the zero point (0 m/s) has been completed.
Perform gain calibration at 2.5 m/s next.
0.0000
Gain calibration starts (the
main display blinks).
A few seconds later
Gain calibration is complete.
0.0000
3
Display the screen shown on the right.
Use the calibrator to input 2.5 m/s.
2.5000
4
Set the cursor at the position of READY and
press the
key to start gain calibration at 2.5
m/s.
When the screen shown on the right is displayed, the gain
calibration at 2.5 m/s has been completed.
Perform gain calibration at 10 m/s in the same way.
2.5000
Gain calibration starts (the
main display blinks).
A few seconds later
Gain calibration is complete.
2.5000