42
USER MANUAL
__________________________________________________________________________________________________________________
18.2 Pieces counting function (PCS)
This function enables to count identical pieces,
e.g. turnbuckles or buttons.
A measurement is performed in two phases:
- first phase - single piece weight calculation
on the basis of defined pieces amount (
5,
10, 20, 50, 100, 200
or
500
pieces),
-
second phase
–
pieces counting.
First phase options:
- PCS . .
–
recalling of a value inserted earlier
(this quantity must be inscribed earlier),
-
PCS SEt
–
set any amount of pieces in a
sample,
-
PCS APW
–
set unitary mass directly,
-
PCS rS
–
inserting number of details in a
sample and receiving of their mass from other
scale connected by RS-232C.
It is advised that single piece weight is not less
than one reading unit and sample weight used
in first phase is bigger than 100 reading units.
To leave function press
MENU
key and then
using
T
key chose
PCS
and
PCS oFF
.
Note:
1.
APW too LOW communicate signalises
that a sample was not put on the pan or if
single piece weight is less than one-tenth
readout plot (counting is not possible).
2.
APW LOW communicate signalizes that
single piece weight is more than one-tenth
but less than one readout plot. (counting
possible but with bigger errors, result
blinks).
3.
In scales equipped with LED display pcs
sign is replaced with “
■
”.
g
5
+10
g