USER MANUAL
__________________________________________________________________________________________________________________
13. Start-up
Plug feeder into ~230V power supply socket.
When the pan is empty plug feeder output
connector into 12V socket at back of the scale.
Autotests and internal calibration will be
performed.
Autotest of balance display.
(autotests of internal electronic elements C1:8
displayed only when any test result is negative)
Showing scale program version.
Internal calibration
–
45 seconds (press key if
You want to terminate calibration)
Ready to work.
Attention:
UnLOAd
communicate means that the
balance is loaded or the transport securing
elements were not clear away.
It is recommended that before you start measuring the internal temperature has
stabilized weight. For this to happen, the weight should remain enabled for at least 2
hours. To maintain the accuracy of the weight it is not recommended to turn off the
power
.
g
0
kg
~45 sekund