18.6 Constant tare memory function (tArE)
This function enables to measure gross weight of a sample placed in a container of a known weight value
(stored in the memory) and to display calculated net weight of the sample. Tare value is recalled from the
memory with
0
or
T
key when the pan is empty. Tare value may be entered using keypad or
by putting container on the pan.
Inscribing tare value to memory:
After pressing
MENU
key and choosing
tArE
function using
T
key, the following options
are available:
-
tAr 0FF
–
leave the function,
-
tAr on
–
activate the function with the
previous tare value,
-
tAr ..
–
sample tare value from the pan,
-
tAr SEt
–
enter tare value with keys:
0
,
,
T
and
MENU
-
out
–
printout a setting value of tare.
Press
T
key when
tAr SEt
is displayed.
By pressing
T
key choose proper
memory cell where tare will be stored:
tAr
01, 02, ... , 10.
Choose inscribing method :
-
MAnUAL
–
inscribing using keys:
0
,
,
T
and MENU,
-
Pan
–
inscribing mass value that is on the
pan.
After storing tare, the scale starts working with
inscribed tare value.
Note:
Tare value is stored in memory also after
unplugging the scale from the mains.
g