To avoid personnel injury or damage to the meter, use only the specified fuse. The specification is 6 3mA
250V fast-melt. 3mA 250V-fuse protection is in the mA current output terminal. To check whether fuses are blown:
•
Turn the rotary switch to
mA
•
Plug the black test into COM, and the red test lead into the mA input.
Use an ohmmeter to check the resistance between test leads .If the resistance is about 20Ω, the fuse is good. An open
reading means that fuse is blown. Replace the fuse as follows.
•
Remove the test leads from the meter and turn the meter OFF.
•
Take off the protector of the meter, remove the four screws by using a standard-blade screwdriver, and then take off
the cover.
•
Replace the blown fuse(s).
•
Reinstall the cover.
4. Specification
4.1. Safety and conformity
Overload protection - V~COM terminal:AC600V/10 seconds; mA terminal:63mA/250V fast-melt fuse
Legal conformity - IEC61010-1(CAT IV 600V�CAT III 1000V�Pollution ClassificationII)
Electromagnetic compatibility - Conforms to IEC61326-1,Group 1�Class B
Surge protection - 8kV(according to IEC61010.1-2001)
Identification Mark - CE
Quality Standard - Develops , designs and manufactures in conformity with ISO 9001
4.2. General feature
Display Screen - digit: display in four digits(5 in current measurement and output)
Refresh display - 2.5 displays/second
Operation temperature and moisture range - 0~40 ºC, relative moisture≤85%(no condensation)
Storage temperature and moisture range - -20 ºC~60 ºC�relative moisture≤90%( no condensation)
Accuracy required temperature and moisture range - 23±5ºC�relative moisture≤75%( no condensation) Temperature
co-efficiency - 0.1 × basic accuracy / °C(temperature range <18°C or >28°C)
Ambient condition for operation - Indoor, outdoor operation(no waterproof),at an altitude of 0~2,000 meter
Indicator for over-range - OL
Continuity/open test - Beeper sounds for resistance readings below threshold, or to indicate a momentary open circuit
Battery type - two AA 1.5V Alkaline (LR6) batteries battery life when using alkaline battery: measures any parameter:
about 100 hours; DC current output(SIMULATE):about 50 hours; DC current output (SOURCE 20mA(1000Ω load):about
2.5 hours.
Low battery - displays battery symbol