background image

Automatic power-off - The default value is 5 mins if no operation which is adjustable.
Preheating time - 10min
Closed – case calibration No internal adjustments needed
Battery access door - Battery or fuse replaceable without voiding calibration
Measurement - 180(L)×90(W)×47(D)mm
Weight - about 500g
Calibration period - one year

4.3. Detailed accuracy indicator

Accuracy is specified for a period of one year after calibration, at 23±5ºC, with relative humidity to 75 %. Accuracy
specifications are given as: ± ( [ % of reading ] + [ number of least significant digits ] ) (note: “number of least significant
digits” means the digits increased or decreased in least significant digits).
Measuring part - Range - Frequency - Accuracy - Remark
DC voltage - 50V - 10mV - 0.5%+4 - • Measuring impedance 10 MΩ (standard value)
• Common mode restrain: 50Hz or 60Hz > 100Db
• String mode restrain:50Hz or 60Hz > 45dB
• Overload protection: 600V (peak –peak value)

AC voltage - 500V - 100mV - 0.5%+20 ( 45Hz~100Hz )/2%+20 ( 100Hz~400Hz )
- • frequency beeps: 45Hz~400Hz
• Adaptive to range from 10% to 100%
• AC conversion: mean value
• Measuring impedence:10 MΩ (standard value)<100pF
• Common mode restrain: 50Hz or 60Hz > 100Db
• Overload protection: 600V(peak –peak value)
Ohm - 5kΩ - 0.001kΩ - 0.5%+4 - • open-circuit voltage: <5V;
• short-circuit current: about 0.1mA.
• leads resistance excluded in the accuracy;
• overload protection: 600V
continuity - 500Ω - 0.1Ω - Short-circuit beep is about 20Ω
• open-circuit voltage: <5V
• short-circuit current: about 0.1mA.
• overload protection: 600V(peak –peak value)
diode - 2V - 0.001V - 1%+20
DC current - 20mA - 0.001mA - 0.2%+4
• overload protection: 63mA/250V fast-melt protection;
• loaded voltage: about 18mV/mA

Содержание AX-C705

Страница 1: ...oduct to see whether it is damaged in the shipment or not Check the materials to see whether they are the same as shown in the packing list Keep the packing materials for late delivery Standard and ch...

Страница 2: ...l understanding of this Manual before use Make sure the requirements in the Manual are strictly followed and keep the Manual for reference whenever needs Mis operation may cause accidents and damages...

Страница 3: ...the level of protection for the pulse compression typical installation location includes Three phase utility power supply equipment for any outdoor power lines or equipment Any outdoor transmission li...

Страница 4: ...n stable 2 3 DC voltage measurement Turn the rotary switch to Insert black probe into COM terminal and insert red probe into _V terminal Connect the probes to the circuit under test and then read the...

Страница 5: ...e electricity fully from all capacitor before testing continuity Turn the rotary switch to D press green button and select test Insert black probe into COM terminal and insert red probe into _V termin...

Страница 6: ...d 250 HART Insert black probe into mA terminal and insert red probe into LOOP terminal Connect the probes to the circuit under test and then read the measured value when stable Press mA button to disp...

Страница 7: ...A output Otherwise low impedance from other position of rotary switch will appear and cause current higher than 35mA to go through the loop circuit Turn the rotary switch to the screen displays OUTPUT...

Страница 8: ...and latch the battery door before using the meter Note The new and old Batteries can not be mixed Max Voltage applied to output terminals about 32V Max current applied to output terminals about 25mA M...

Страница 9: ...to IEC61326 1 Group 1 Class B Surge protection 8kV according to IEC61010 1 2001 Identification Mark CE Quality Standard Develops designs and manufactures in conformity with ISO 9001 4 2 General featu...

Страница 10: ...ark DC voltage 50V 10mV 0 5 4 Measuring impedance 10 M standard value Common mode restrain 50Hz or 60Hz 100Db String mode restrain 50Hz or 60Hz 45dB Overload protection 600V peak peak value AC voltage...

Страница 11: ...transmitter 20mA 0 000 22 000mA 0 001mA 0 2 4 external power supply 5 28V Max load 1k at 20mA LOOP Loop circuit power 24V 10 Max output current 25mA Max Voltage applied to output terminals about 32V...