Do not apply any voltage to output terminals except in SIMULATE mode. Any wrong application may damage internal
circuit.

2.11. Constant current output (SOURCE mode)

Turn the rotary switch to “”; the screen displays “OUTPUT” and “LOOP POWER”; output is set as 0mA.

Insert black probe into “mA”” terminal ,and insert red probe into “+LOOP” terminal;

Press ”mA%” button to display current value in percentage form.

Press <LEFT>, <RIGHT>, <UP>, <DOWN> buttons to set output value.

Connect the leading wire to the circuit under test.

2.12. Constant current output (SIMULATE mode)

SIMULATE mode refers to simulate a group of loop-circuit current transmitter with the Meter. If there are cases of an
external DC voltage (range from 5-28 V) and tested loop current in series, then select SIMULATE mode.
Warning
Before connecting the testing lead to loop circuit current, turn the rotary switch to any position of mA output. Otherwise,
low impedance from other position of rotary switch will appear and cause current higher than 35mA to go through the
loop circuit.

Turn the rotary switch to “”; the screen displays “OUTPUT” and “SIMULATE”; output is set as 0mA.

Insert black probe into “mA”” terminal ,and insert red probe into “mA” terminal; Press ”mA%” button to display
current value in percentage form.

Press <LEFT>, <RIGHT>, <UP>, <DOWN> buttons to set output value.

Connect the leading wire to the circuit under test.

2.13. Manual-step output

In constant current output mode, press green button to select manual-step output function: increases or
decreases in 25% or 100% step. Press <UP>, <DOWN> button may increase or decrease current in 25% or 100% step.

2.14. Setup function

Setup function may:

Start and close automatic power-off function.

Open and close buzzer.

In power-off state, press button and turn the rotary switch from OFF to any other position simultaneously. Wait for 2
seconds and then loosen. Press button to select setup function; press button to change setup state; press green button to
hold setup value.

Содержание AX-C705

Страница 1: ...oduct to see whether it is damaged in the shipment or not Check the materials to see whether they are the same as shown in the packing list Keep the packing materials for late delivery Standard and ch...

Страница 2: ...l understanding of this Manual before use Make sure the requirements in the Manual are strictly followed and keep the Manual for reference whenever needs Mis operation may cause accidents and damages...

Страница 3: ...the level of protection for the pulse compression typical installation location includes Three phase utility power supply equipment for any outdoor power lines or equipment Any outdoor transmission li...

Страница 4: ...n stable 2 3 DC voltage measurement Turn the rotary switch to Insert black probe into COM terminal and insert red probe into _V terminal Connect the probes to the circuit under test and then read the...

Страница 5: ...e electricity fully from all capacitor before testing continuity Turn the rotary switch to D press green button and select test Insert black probe into COM terminal and insert red probe into _V termin...

Страница 6: ...d 250 HART Insert black probe into mA terminal and insert red probe into LOOP terminal Connect the probes to the circuit under test and then read the measured value when stable Press mA button to disp...

Страница 7: ...A output Otherwise low impedance from other position of rotary switch will appear and cause current higher than 35mA to go through the loop circuit Turn the rotary switch to the screen displays OUTPUT...

Страница 8: ...and latch the battery door before using the meter Note The new and old Batteries can not be mixed Max Voltage applied to output terminals about 32V Max current applied to output terminals about 25mA M...

Страница 9: ...to IEC61326 1 Group 1 Class B Surge protection 8kV according to IEC61010 1 2001 Identification Mark CE Quality Standard Develops designs and manufactures in conformity with ISO 9001 4 2 General featu...

Страница 10: ...ark DC voltage 50V 10mV 0 5 4 Measuring impedance 10 M standard value Common mode restrain 50Hz or 60Hz 100Db String mode restrain 50Hz or 60Hz 45dB Overload protection 600V peak peak value AC voltage...

Страница 11: ...transmitter 20mA 0 000 22 000mA 0 001mA 0 2 4 external power supply 5 28V Max load 1k at 20mA LOOP Loop circuit power 24V 10 Max output current 25mA Max Voltage applied to output terminals about 32V...

Отзывы: