AX-C705
1. Introduction
1.1. Generalization
The Meter is a battery-operated process maintenace field device which combined both digital-parameter function with
process-signal source together.
Functioning as follows:
•
Measurement function of TC voltage, DC voltage , Ohms, DC, Continuity and Diode; Data reserving function;
•
Max. Voltage applied to output terminals: about 32V; Max. current applied to output terminals: about 25mA.
•
Output function of DC (Constant, Manual-Step and SIMULATE modes)
•
Loop-circuit testing function; Powering the loop circuit with 24 Voltage and testing the current; Built-in 250Ω HART
loop resistance
•
Safety: complied with IEC 61010-1 safety requirements for measurement, control and electronic instruments for lab
use.
1.2. Opening the Meter Case
Check the product to see whether it is damaged in the shipment or not. Check the materials to see whether they are the
same as shown in the packing list. Keep the packing materials for late delivery.
Standard and chosen accessories supplied are listed as follows. Chosen ones are bought at purchaser’s options.
Standard accessories:
•
1x testing wire(with alligator clips)
•
1x user’s manual
•
2x 1.5V (LR6)Alkaline batteries
•
2x 63mA/250V Fast Fuses
1.3. Safety Information
The design, manufacture and test of the Meter reach the IEC 61010-1 Safety Requirements. This Manual contains all
warnings and safety regulations that must be followed to ensure safe operation and retain the meter in safe condition.
Read the following instructions before operation.
Mark
in the Meter means the operator needs to make reference to related parts in the Manual to ensure safe operation.
Warning - indentifies conditions and actions that pose hazard(s) to the user; Caution - indentifies conditions and actions