Axetris AG
LGD Compact Integration Guide V1.8
04.12.2019
24/34
Instructions for cable packaging
The LGD compact module will be delivered default with this electrical cable from sensor housing to electronic
boards:
All EMC tests were done with this cable packaging. If this cable cannot be used due to integration issues
please be aware of the following recommendations for the cable packaging:
•
Conductor cross section 17x 0.14 mm² (signal line), AWG 26
•
Fully shielded cable with good shield connection
•
Max. cable length = 350mm (in general as short as possible)
•
Twisted wire couples according to the following diagram:
Figure 19
Diagram of the cable wiring
Figure 20
Pin connection of the LGD compact sensor module