Axetris AG
LGD Compact Integration Guide V1.8
04.12.2019
31/34
27
– 30
00 00 C8 42
Analog output: Max concentration
(here: 100)
31
– 34
00 00 00 00
Analog output: Offset
(here: 0)
35
– 38
0A D7 23 3C
Analog output: Span
(here 0.01)
41
– 42
00 00 20 41
Integration time
(here: 5)
43
– 46
00 80 BB 44
Concentration gas 2 Offset
(here 1500)
47
– 50
33 33 73 3F
Concentration gas 2 Span
(here 0.95)
51
– 54
00 00 00 00
Diagnostic Mode (1 = active / 0 = inactive)
(here 0)
For further details about Diagnostic mode, contact Axetris AG.
55
– 58
00 00 00 00
Laboratory Mode (1 = active / 0 = inactive)
(here 0)
For further details about Laboratory mode, contact Axetris AG.
59
– 62
00 00 00 40
Analog switch
(here 2)
63
– 66
00 00 00 00
Reserved variable for future use (dummy)
67
– 70
00 00 00 00
Reserved variable for future use (dummy)
71
– 74
00 00 00 00
Reserved variable for future use (dummy)
75
– 78
00 00 00 00
Trailing zeros
79
– 80
7F
7D
End of packet with checksum
Please note that this general description may be subject to changes depending on single
product execution.
Configure user data
The
Config
command is used to change the user data of the system. The parameter of this command is the
list of the user data.
Table 15
Configure user data command with description of the bytes
Byte
Possible HEX
Description
1
– 4
7B 43 50 00
Start of packet with size (80 bytes)
5
– 6
00 EA
Additional command
7
– 10
00 00 40 40
Concentration gas 1 Offset
(here 3)
11
– 14
CD CC 8C 3F
Concentration gas 1 Span
(here 1.1)
15
– 18
00 00 00 00
Reserved variable for future use (dummy)
19
– 22
00 00 00 00
Reserved variable for future use (dummy)
23
– 26
00 00 00 00
Analog output: Min concentration
(here: 0)
27
– 30
00 00 C8 42
Analog output: Max concentration
(here: 100)
31
– 34
00 00 00 00
Analog output: Offset
(here: -1)
35
– 38
0A D7 23 3C
Analog output: Span
(here 0.01)
41
– 42
00 00 20 41
Integration time
(here: 10)
43
– 46
00 80 BB 44
Concentration gas 2 Offset
(here 1500)
47
– 50
33 33 73 3F
Concentration gas 2 Span
(here 0.95)